Several common problems and solutions of SG series water bath sterilization cabinet
Date: 2025-07-02Read: 28
The SG series water bath sterilization cabinet adopts computer for full process monitoring. Various parameters can be modified, recorded in real-time, and printed out, greatly improving the reliability and intelligence of the entire machine, making it better suited for the needs of modern industrial production in enterprises.
The common problems and solutions of SG series water bath sterilization cabinets are as follows:
  1. No response after power on
Power inspection: Confirm that the power socket is powered on and in good contact, and check if the power cord is broken or if the fuse is blown. ‌
Circuit testing: Check if the transformer is burnt out or if the lead welding is loose, and use a multimeter to test the continuity of the circuit. ‌
  2. Abnormal heating
Temperature control device: Check if the temperature controller is damaged or installed in the wrong position (such as installing the upper and lower layers upside down), and recalibrate or replace it. ‌
Heating tube malfunction: If there is no heating or the temperature cannot reach 121 ℃ after starting, it may be a damaged heating tube and requires professional maintenance. ‌
  3. Uneven temperature
Steam supply: Ensure stable steam pressure, check pipeline sealing to prevent leakage. ‌
Loading adjustment: Avoid stacking too many utensils or placing them close to the cabinet door to ensure hot air circulation. ‌
  4. Equipment leakage
Sealing inspection: Focus on inspecting the door seal and drainage valve. If water leakage is found, replace the sealing ring or clean the blockage of the drainage hole. ‌
Pipeline maintenance: Regularly clean scale and impurities to prevent pipeline blockage and abnormal pressure. ‌
  5. Program exception interrupt
Water shortage protection: Check if the water level meets the standard (170cm standard water level), refill the water and restart the program. ‌
Circuit maintenance: Check the contact condition of the relay and replace faulty components if necessary. ‌
  6. Ozone/UV function failure
Component inspection: Confirm that the door is tightly closed, check if the ozone tube and ultraviolet lamp are damaged, and replace accessories if necessary. ‌
Circuit maintenance: Check whether there are short circuits or burn marks on the circuit board, and professional personnel need to repair them. ‌
  7. Maintenance suggestions
Regular cleaning: Clean the drainage holes and residual materials in the inner container to prevent bacterial growth. ‌
Safe operation: Power must be cut off during maintenance to avoid the risk of electric shock. ‌