Bottles are ubiquitous in daily life, and with the popularity of automatic filling lines, the verticality of bottles is becoming increasingly high. Bottles with poor verticality may cause the nozzle to fail to insert into the bottle during filling, knocking over the bottle and affecting the production line. It may also cause the liquid to be filled to spray outside and damage the filling line; Inverted bottles can get stuck in the filling line, causing the production line to shut down or the machine to be damaged. Therefore, it is necessary to implement verticality testing on bottles.
Sai Cheng has successfully developed an electronic axis deviation measuring instrument called ZPY-G electronic deviation tester, which meets the new national GMP standard, by testing the physical property index of vertical axis deviation of cosmetic bottles, medicinal glass bottles, wine bottles, plastic bottles, and other bottles.
In terms of appearance design, the new ZPY-G electronic axis deviation measuring instrument from Saicheng R&D fully embodies humanization. Greatly reducing the workload of operators and the errors caused by human error, an automatic start arm has been added, which means that users can easily achieve the desired height by clicking on the software's preset "up", "down", and "stop" buttons to adjust the measuring rod, instead of manually adjusting it like mechanical methods, which consumes the energy of staff and wastes the time of sample testing.
Operation Method
Fix the bottom of the bottle onto the rotating disc of the horizontal plate, so that the bottle mouth contacts the dial indicator. Rotate 360 ° to read the larger and smaller values, and half of the difference between the two is the vertical axis deviation value. The instrument cleverly utilizes the high concentricity of the four jaw self centering chuck, combined with a set of high degree of freedom brackets that can freely adjust height and orientation, to meet various glass bottle inspections.
The Saicheng ZPY-G electronic axis deviation tester is suitable for detecting the vertical axis deviation of bottle containers in industries such as ampoules, penicillin bottles, pharmaceuticals, food, cosmetics, etc. This instrument complies with multiple national and national standards: YBB00332002 - 2015 、 YBB00332003 - 2015 、YBB00032004-2015、GB/T 84522008、GB-2639、QB 2357、QB/T 1868。 According to standards such as YBB0033202-2015 "Low Borosilicate Glass Ampoule", innovative design has been carried out, and the fully automated testing process has greatly improved the accuracy of testing. It is an instrument used by pharmaceutical companies and drug testing institutions.