Maintenance of components for laboratory mobile suction hood
Date: 2025-09-05Read: 10

Laboratory mobile suction hoodThe series is based on the product concept of compliance, environmental protection, health, and safety. It is equipped with activated carbon molecular filters and air particle filters, which can adsorb most chemical gases. When researchers conduct chemical experiments in the cabinet, harmful gases are adsorbed by the filters in the filtering device, and the filtered air circulates indoors. The clean air type fume hood does not exhaust to the outside, thus reducing the operating energy consumption (cost) of air conditioning and ventilation systems; Moreover, due to the adsorption of harmful gases by filters, they are not directly discharged outdoors, which can reduce pollution to the atmospheric environment.

Characteristic description:
1. Metal components: The main material is high-quality cold-rolled steel plate with a thickness of ≥ 1.2mm, coated with electrostatic powder coating and durable chemical resistant lead-free coating to maintain high smoothness and minimize the effects of corrosion and moisture.
2. 4.3-inch LCD touch screen display, real-time temperature and humidity environment monitoring, fan monitoring, VOC concentration environment monitoring, and integrated alarm system.
3. Equipped with high-quality centrifugal fans, 24V current, stable performance, and no sparks or static electricity.
4. Interface: Anti electromagnetic interference power cord.
5. Working principle: The device is equipped with a built-in filtering module, which collects and filters hazardous exhaust gases from the emission source. The filtered clean air is directly discharged into the room.
Functional component maintenance:
Filtration system - The ductless activated carbon filter element is replaced every 3 months, and the HEPA filter is washed with water and dried.
Spray electronic cleaner when the air flow control valve knob is stuck, calibrate the relationship between the opening and the air volume.
The mobile base pulley groove is cleaned of debris every month, and lithium based grease is added to the bearings to prevent rust.
Maintenance process:
Annual comprehensive maintenance:
Remove all joint bolts, apply corrosion-resistant thread sealant and re tighten.
Detect the cracking of the airway (light transmission method), replace it if it ages and becomes hard.
Performance calibration:
Measure the wind speed at the hood mouth with an anemometer, and calibrate the sensor if the deviation from the displayed value of the controller is greater than 10%.