The main characteristics and usage of the six link intelligent constant temperature magnetic stirrer
Date: 2024-05-24Read: 4
  Six link intelligent constant temperature magnetic stirrer HCJB-6SProduct Features
  1.The 30 ° inclined control panel is suitable for both sitting and standing perspectives.
  2.Using aluminum plate heating, it has the characteristics of large heating area, uniform heating, and fast heating.
3. Equipped with an over temperature alarm function.
4. Adopting a DC brushless motor, it has stable performance, low noise, long service life, and no spark generation.
5. Digital speed display function.
  6.The temperature control adopts fuzzy PID control algorithm, LCD single screen display, self-tuning function, with high measurement accuracy, small temperature surge, single key touch operation, internal and external thermocouple temperature measurement, thyristor control output, 160-240V wide voltage power supply, and has a break protection function.
7. Timer function, the timer function can choose to run timer or constant temperature timer.
8. Smooth at low speeds, strong at high speeds.
  9.Heating and stirring can be performed on 50-2000ml standard or non-standard reaction bottles.
10. Control heating separately stir. Timed.
  Six link intelligent constant temperature magnetic stirrer HCJB-6SUsage:
1. Fix the upright pole in the screw hole above and behind the mixer, adjust the height of the cross clamp, use a universal clamp to fix the reaction bottle, put in a suitable mixer, insert the power supply~220 ∨, and turn on the switch.
2. Press the temperature button to increase or decrease the temperature required for adjustment, and then press the temperature button to start heating (display A as external sensor, B as internal sensor)
3. Press the stirring button to start stirring, and press the speed button to increase or decrease the adjustment speed.
4. Press the time button to increase or decrease the time required for adjustment (the time is counted down and starts counting when the set temperature is reached)
5. The appliance is equipped with a break protection function. When the thermocouple connection is poor, the display window shows E-1 as an external sensor fault, E - as an internal sensor fault, and E-3 as an ambient temperature fault. The appliance stops heating and needs to be checked before use.
When the temperature deviates too much from the set value, the self-tuning function can be activated for adjustment. Please refer to the operating instructions for the operating steps.