Analysis of Installation Conditions for Drug Stability Test Chamber
Date: 2020-12-16Read: 9

The drug stability test box is used in the pharmaceutical industry to conduct tests on factors affecting drug stability and determine the factors affecting drug quality. The product meets the requirements of ICH, FDA, and current pharmacopoeia for strong light exposure testing. Creating a long-term stable temperature for drug failure assessment is a scientific approach. The humid and light environment is suitable for pharmaceutical companies to accelerate drug and new drug testing, long-term testing, and high-temperature testing. Conducting drug stability tests is a good choice for pharmaceutical companies.

Analysis of installation conditions for drug stability test chamber:
Before installing the drug stability test chamber, we need to find a suitable installation location to ensure that the equipment does not vibrate due to uneven ground, and does not age due to excessive dust and poor ventilation in the area.
2. The power supply connected to the device should be stable at 380 volts. If tested for a long time, the aging rate of internal components of the device will accelerate.
When installing the test chamber, it should be at least 80 cm away from the wall, otherwise it will affect the maintenance of the equipment and may fail due to heat dissipation.
If there are any difficult problems during the installation of the test chamber, please consult the manufacturer in a timely manner to avoid equipment failure caused by unauthorized operation.
How to check and use drug stability test kits?
1. Check if the power supply is in good condition, check the voltage requirements for power supply usage, etc., to avoid problems after powering on.
2. Check if the appearance of the test chamber is damaged, open the instrument to see if there is any abnormal noise, and try pressing the knob or button of the instrument to see if it is flexible and not loose.
3. When using the device, it can be powered on and preheated for 15 seconds to ensure normal operation after startup.
4. It is necessary to set an over temperature protection temperature, which is generally 10-20 degrees Celsius higher than the working temperature.
The drug stability test chamber is used for rigorous research in the pharmaceutical industry, pharmaceuticals, biotechnology, food industry, electronics industry, and all related industries including life sciences. The World Health Organization principle requires long-term stability testing to be conducted under conditions of 25/60% relative humidity. In the accelerated test, a humidity check of 40/75% RH is conducted for 6 months, which belongs to the field of stability testing systems in the pharmaceutical industry. It mainly simulates temperature, humidity, and light tests under environmental climate conditions.