model |
3321 |
|
detection technology |
Using a separate high-speed processor to detect the flight time of individual particles in accelerated airflow |
|
particle size range |
0.5-20 µ m (aerodynamic diameter) 0.37-20 µ m (light scattering diameter) |
|
Aerodynamic diameter resolution rate |
0.02 µ m (1 µ m), 0.03 µ m (10 µ m) |
|
resolution |
Particle size resolution |
32 channels/10 times particle size. There are a total of 52 channels. 1024 bit raw flight time data in unrelated mode (4ns/bit) |
Resolution of light scattering intensity |
16 channel light scattering intensity. 64 channel raw light scattering intensity data |
|
Detecting particle types |
Suspended particulate matter and non-volatile liquids |
|
Maximum particle detection concentration |
1000 pieces/cm3 at 0.5 µ m. The probability of overlapping events occurring is less than 5%; 1000 pieces/cm3 at 10 µ m. The probability of overlapping events occurring is less than 10%: the upper limit of available data is 10000/cm3 |
|
Minimum particle detection concentration |
0.001 pieces/cm3 |
|
Concentration range |
± 10% count value plus statistical count deviation |
|
Maximum aerodynamic particle size processing speed |
>200000 pieces/s |
|
Total sampling flow |
5.0L/min |
|
Sheath gas flow rate |
4.0L/min |
|
Sample gas flow rate |
1.0L/min |
|
Operating temperature |
10-40 degrees |
|
Atmospheric pressure correction |
Automatic calibration within 400 to 1030mbar (complete calibration within 700-1030mbar) |
|
laser source |
30mV.655nm diode laser |
|
detector |
Avalanche type photomultiplier tube (APD) |
|
Front display panel |
320 x 240 pixels |
|
Operating Temperature |
10-40 degrees |
|
Operating Humidity |
10-90% |
|
power supply voltage |
110-240VAC. 50/60Hz. 100W or 24VDC |
|
contact method |
RS-232 (9-pin) interface |
|
output method |
Digital I/O |
|
Configure analog output |
BNC(0-10V) |
|
Pulse simulation output |
BNC |
|
Flight time digital output |
BNC external controller uses a 15 pin interface (3 inputs, 3 outputs). 2 analog input interfaces |
|
Overall dimensions |
Aerosol injection port |
Outer diameter 3/4 inch |
outer box dimensions |
38 x 30 x 18cm |
|
weight |
10kg |
|
Aerodynamic particle size spectrometer
Model: 3321
TSI in the United StatesThe 3321 Aerodynamic Particle Size Spectrometer (APS3321) is a high-performance, multi-purpose next-generation particle measurement instrument with a unique design that can simultaneously measure the following two properties:
Aerodynamic particle size: APS 3321 can measure the aerodynamic particle size of particles in real-time through precise time-of-flight (TOF) technology. Its particle size measurement range is 0.5-20 µ m. Due to the fact that aerodynamic particle size counting based on flight time is only related to particle shape. Thus avoiding refractive index and meter scattering instruments. In addition, the monotonic curve of time-of-flight particle size measurement ensures high resolution throughout the entire particle size measurement range.
Optical scattering intensity: measured through optical scattering techniques. The measured particle size of TSI APS 3321 in the United States can reach 0.37-20 µ m. Although the light scattering intensity is not always a reliable characterization parameter for particle size. But it is still a parameter of interest to people. APS 3321 can provide data on the aerodynamic particle size and optical scattering intensity of the same particle separately. And stored separately.
By measuring the aerodynamic particle size and optical scattering intensity of each particle through the same sensor, you can gain exciting new insights into the composition of aerosols. APS 3321 uses a continuous bimodal optical system with US patent * technology to detect the frequency of particle overlap events (multiple particles simultaneously passing through the detection area). And identify weak signals close to the detection limit. Therefore, the obtained detection results are solid, reliable, and high-quality data
APS 3321 has a beautiful and elegant appearance, as well as a simple and easy to operate front panel (including control knobs and built-in display screen). The control knob facilitates users to scan and view data on the display screen. Different functions of monitoring or controlling instruments. In addition, the instrument also has a volume flow controller controlled by microprocessor, atmospheric pressure correction, independent sheath gas sampling pump, and total sample gas sampling pump. These features ensure that APS 3321 can operate normally and maintain automatic calibration under various conditions. The Particle Instrument Management Software (AIM) is a sampling software based on a 32-bit WINDOWS program, which ensures that each 3321 has complete system and data control functions.
application
The APS 3321 aerodynamic particle size spectrometer measures the particle size distribution and has a wide range of applications.
including:
Respiratory toxicity study
Drug delivery research
Biological toxicity testing
atmospheric science
Environmental air monitoring
indoor air quality monitoring
Membrane and cleanroom testing
Study on the characteristics of aerosols used for calibrating particle instruments
Spray measurement technology
Performance evaluation based on aerodynamic particle size measurement instrument
Powder particle size distribution
basic research
Remarks: TSI3321 from the United States, TSI Aerodynamic Particle Sizer 3321 from the United States, Acting TSI3321 from the United States, TSI Particle Sizer 3321 from the United States, Acting Particle Sizer 3321 from Zhongyi Environmental Science, TSI Laser Particle Counter from the United States, TSI Aerosol Generator from the United States