Product Introduction
More than 260 measurement programs are pre-set, which can use Orion AQUAFast, Merck, CHEMetrics or Hach reagents and various sizes of colorimetric cells (bottles) to easily and accurately measure the following parameters:
COD、BOD、TOC、UV254、SUVA、 Chlorophyll a, pH, ammonia, ammonium, dissolved oxygen, acidity, alkalinity, hardness, suspended solids SS, chromaticity, silicon, hydrazine, phosphate, oxygen scavenger, chlorine, chlorine dioxide, monochloramine, hydrogen peroxide, cyanuric acid, ozone, boron, bromine, iodine, nitrogen, oxygen, fluoride, chloride, sulfide, chromate, molybdate, nitrate, nitrite, sulfate, sulfite, lead, chromium, cadmium, mercury, arsenic, manganese, zinc, copper, iron, molybdenum, nickel, antimony, tin, gold, silver, platinum, palladium, potassium, sodium, calcium, magnesium, aluminum AOX、 Volatile organic acids, surfactants, cyanide DEHA、 Detergent, ethanol, formaldehyde, phenol.
Scientific research, routine testing
The combination of high-intensity xenon lamp (emitting flashing light during measurement) and dual beam optical system enables AquaMate 8000 to obtain excellent data in the full ultraviolet to near-infrared spectral range.
Dual beam optical system
Due to the strong light emitted by xenon lamps, even if a spectrometer is used to distribute a small portion of the light to the built-in reference detector, it will not affect the sample testing. This allows for reference beam calibration to be performed simultaneously with each sample measurement, avoiding errors caused by the light source. Due to the effective correction of measurement errors caused by changes in light source intensity and detector sensitivity using reference beams, the accuracy of each measurement result can be maximized!
Ready to use - maintenance free xenon lamp
The AquaMate 8000 uses a xenon lamp as the light source, providing perfect performance with a wavelength range of 190-1100nm. At the same time, in the ultraviolet range, the high-intensity light source provides strong protection for applications such as life sciences, environmental testing, and organic chemistry. Xenon lamp light source does not require preheating, it can be turned on and tested immediately. Because the sample will not be continuously irradiated, it will not affect sensitive samples. And it will not change the temperature inside the sample chamber, ensuring the stability of the sample temperature. Because xenon lamps only light up during measurement, they can usually be used for 5-7 years and do not require special maintenance. Compared with traditional light sources, this saves users about 30000 yuan in maintenance costs within 5 years.
Excellent 1.8nm bandwidth
The AquaMate 8000 has improved sensitivity and optimized resolution to the greatest extent possible. Because a bandwidth of 1.8nm allows more light energy to reach the sample, resulting in lower detection limits and better signal-to-noise ratio.
Fast wavelength scanning
The AquaMate 8000 UV/Vis water quality analyzer is equipped with wavelength scanning technology, which can quickly obtain high-quality spectral data. It has increased the scanning speed to 4200nm/min. With an ultra wide range of luminosity, even small changes in absorbance can be accurately measured, even for strongly absorbing blank samples. Scanning data can be analyzed to determine the wavelength values of peaks and valleys, and peak height and 3-point calculations can be used to obtain a tilted baseline. The scanned data can be saved to a USB drive.
Color pool fixed seat
The AquaMate water quality analyzer can provide the following three types of mounting brackets, which can be flexibly matched with various colorimetric pools (bottles). 13-25mm adjustable fixed bracket, suitable for circular colorimetric bottles;
20/50mm rectangular colorimetric pool fixing bracket; 10mm square colorimetric cell fixing bracket. For users with a large sample size, they can also choose a 6-cell rack, which can simultaneously place 6 10mm colorimetric cells.
optical design |
Dual beam, built-in reference detector |
spectral bandwidth |
1.8nm |
Light source (standard lifespan) |
Flashing xenon lamp (5 years) |
detector |
Double silicon photodiode |
wavelength range |
190 - 1100nm |
spectrum |
|
stray light |
<0.08% T at 220nm and<0.1% T at 340nm |
display |
Backlit LCD, 9.7 * 7.1cm |
keyboard |
Sealed film button |
connect |
Front A-type USB port - USB flash drive, |
size |
30W*40D*25H cm |
weight |
8.6kg |
power supply |
100 - 240V,50 - 60Hz |
Order Number |
description |
AQ8000 |
AquaMate8000 UV visible spectrophotometer, pre stored measurement program USB flash drive, 6-connected cell, 3-digit cell holder, 1-bit cell holder, user manual, dust cover, USB cable, 110-220V power supply |
AQ006C |
6 consecutive pools |
AQ003C |
3 pool racks |
AQ001C |
1 pool rack |
AQ0USB |
Pre stored measurement program USB flash drive |
AQCHNV |
Power cord‐ Chinese style plug |
AQ240V |
Power cord‐ 240V British plug |
AQ220V |
Power cord‐ 220V European style plug |
AQ110V |
Power cord‐ 110V American plug |
AC2V24 |
24mm color comparison bottle, pack of 12 |