Welcome Customer !

Membership

Help

Beijing Jingke Zhichuang Technology Development Co., Ltd
Custom manufacturer

Main Products:

pharmamach>Products

Beijing Jingke Zhichuang Technology Development Co., Ltd

  • E-mail

    bjjkzc01@163.com

  • Phone

    15810615463

  • Address

    Mapo Xiangyue Sijisan District, Shunyi District, Beijing

Contact Now

Beijing Professional High Temperature Thermistor Material Parameter Measurement and Analysis System

NegotiableUpdate on 07/29
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Description: PTC/NTC-500 High Temperature Thermistor Material Parameter Measurement and Analysis System Keywords: Thermistor Material, Resistance, Volt Ampere, Voltage

Product Details

PTC/NTC-500 high temperatureThermistor Material Parameter Measurement and Analysis System

keyword:heat-sensitive material,resistance,Volt ampere, voltage

Product background knowledge:

PTC stands for Positive Temperature Coefficient, which means positive temperature coefficient. It generally refers to semiconductor materials or components with a high temperature coefficient. The PTC we usually refer to is a temperature coefficient thermistor, abbreviated as PTC thermistor. PTC thermistor is a typical semiconductor resistor with temperature sensitivity. When it exceeds a certain temperature (Curie temperature), its resistance value increases step by step with the increase of temperature. Ceramic PTC is a semiconductor ceramic mainly composed of barium titanate (or strontium, lead), with a small amount of donor (Y, Nb, Bi, Sb), acceptor (Mn, Fe) elements, and glass (silicon oxide, aluminum oxide) additives added, and sintered. Ceramic PTC has low resistance below the Curie temperature.

1、 Product Overview:

The PTC/PTRC-500 high-temperature thermistor material parameter measurement and analysis system is capable of testing different types of PTC samples. Under different voltage and temperature conditions, it can obtain the resistance temperature characteristic curve, volt ampere characteristic curve, and resistance voltage (voltage effect curve) of different types of samples. It is possible to analyze the curves of the resistance and resistivity of the tested sample as a function of temperature and time, as well as the volt ampere and current time characteristics at different temperatures.

2、 System features:

1. Real time display of automatic boost, temperature, resistance, voltage and other related values in the testing system

2. The test result curve visually displays R-T, V-I, I-T.

3. Electric heating technology, intelligent control of temperature rise, adjustable heating rate

4. Infrared temperature measurement technology, real-time collection of PTC material temperature

5. The software control system can print or export electronic data of the test results

6. Automatic sample lifting bracket, feeding the placed sample into the heating chamber

2、 Technical parameters

1. Test temperature: 0-600 ℃ (standard type), 600-800 ℃, 800 ℃ -1200 ℃ can be customized.

2. Resistance range: 100 Ω -10 18 Ω

3. Voltage: 0- ± 600V

4. Temperature control accuracy: ± 2 ℃

5. Measurement accuracy: ± 1 ℃

6. Electrical resistivity: 10-10 15 Ω

7. Collection frequency: 1ms

8. Collection method: automatic real-time

9. Electrode Material: Platinum Iridium Electrode

10. Measurement environment: inert atmosphere, reducing atmosphere, vacuum atmosphere

11. Measurement method: Two line method

12. Test channel: single channel or dual channel

13. Sample size: φ < 25mm, d < 4mm

14. Electrode material: Platinum iridium alloy electrode (high temperature resistance, oxidation resistance)

15. Insulation material: 99 alumina ceramic

16. Data storage format: Automatically analyze data, classify and save, combine samples and measurement plans to generate experimental plans required by the system, and output TXT, XLS, BMP and other format files

17. Data transfer: USB

18. Compliant with standard: GB-T 10581-2006

19. Power supply: 220V ± 10%, 50Hz

20. Working temperature: 5 ℃ to+40 ℃;

21. Working humidity: At+40 ℃, the relative humidity can reach up to 95% (without condensation)

22. Equipment size: L360mm * W370mm * H510mm

23. Weight: 22kg