Carbon dioxide isotope analyzeryesLGR(Currently under ABB Canada)to2010Products launched in the year, capable of measuring simultaneouslyδ13C, δ18Oand12CO2,13CO2, CO18O, CO2as well asH2OConcentration. It can be measured continuously in situ or manually injected intermittently, and is insensitive to interference from hydrocarbons such as methane,δ13C, δ18OThe data is not affected by rapid changes in carbon dioxide concentration in the field, making it suitable for laboratory and field use and meeting various requirements for various purposesCO2Isotope measurement equipment. Recently, many researchers have developed this deviceCombining with traditional equipment to develop many new measurement methods has become a promising new direction.
characteristic:
1. Simultaneous measurementδ13C,δ18O,CO2Concentration andH2Oconcentration
2. Wide range,CO2The concentration range is380~25000 ppm
3. 1 HzHigh frequency measurement
4. Ultra stable measurement, high precision and accuracy, minimal drift
5. Combination of in-situ continuous measurement and manual intermittent injection
6. Synchronous measurementCO2Isotope molecules, isotope ratios are not affectedCO2The impact of rapid changes in concentration
Performance indicators:
Repeatability/accuracy(1σ, onesecond/ 10second/ 100second):
δ13C: one point five‰ / 0.5‰/ 0.15‰
δ18O: eight‰ /2.5‰ / 1‰
[12CO2]: zero point fourppm / 0.15 ppm/ 0.05 ppm
[H2O]:100 ppm (60 seconds)/50 ppm (300 seconds)
Measurement speed:
All parameter measurement frequencies can reach1 Hz
Measurement range (full range linear, meeting all technical specifications):
CO2:380 ~ 25000 ppm
H2O:4000 ~ 60000 ppm
Optional measurement range:
CO2:0 ~ 50000 ppm
H2O:0 ~ 70000 ppm, without condensation
Response time:
1Second (flow rate response)<>Optional external pump needs to be configured in seconds)
Accuracy:
Better than 1% of the reading (without calibration)
Drift(1Hourly average, standard temperature and pressure,24Hour):
δ13C: <>
Environmental conditions:
Sample temperature:-20~50℃
Operating temperature:0~45℃
ambient humidity:0~100% RH, without condensation
Temperature control accuracy:
0.003 ℃
Pressure control accuracy:
0.001 torr
output:
number(RS 232)EthernetUSB
Electricity demand:
115/230 VAC,50/60 Hz, 66 W (portable),350 W
Size and weight:
28 cm(H)x 97 cm(W)x 56 cm(D), 68 kg (desktop)
40 cm(H)x 48 cm(W)x 61 cm(D), 50 kg (rack mounted)
18 cm(H)x 47 cm(W)x 36 cm(D), 17 kg (portable)
Ordering information:
Model number(Model):
GLA431-CCIA2 (desktop) GLA331-CCIA2(Rack)GLA132-CCIA2 (portable)
Optional attachments:
1. MIU: 16/8-channel multiplexer
2. SC-1000:16/8-channel synchronous multiplexer (produced by LICA)
3. ACC-DP40: N940 vacuum pump (gas update time 1 second)
4. DCS-200: Dynamic dilution system, capable of automatic dilution and expanding the range by 100 times
5. OPT-BATCH-Injection: Intermittent injection system (integrated with analyzer) that can be manually injected to measure samples inside the airbag
6. LI-6000: Intelligent injection controller, connected to TOC and CCIA, for solid or liquid C isotope determination
manufacturer:
ABB Canada