LGR (currently under ABB Canada)Carbon monoxide/Simultaneous measurement of environmental levels using a carbon dioxide analyzerCO、CO2andH2OConcentration, applied in trace gas monitoring, vorticity related flux measurement, box method flux measurement, exhaust emission measurement and other fields. The analyzer is particularly suitable for field measurements and has been successfully installed inNASAofDC-8Aircraft used for measuring the atmosphere in the troposphere and stratosphere. The analyzer adoptsLGRBuilt in temperature control technology provides higher precision, accuracy, and smaller drift.
CO/CO2The analyzer adoptsLGROutput spectrum of off-axis integrating cavity with patented design(OA-ICOS)Technology (patent number: 6795190), which eliminatesCRDSDuring the measurement process, technology requires continuous matching of the optical cavity and laser wavelength to improve the weak signal intensity, eliminating the need for complex laser alignment adjustment, temperature control, and wavelength monitoring in the analyzer. Real time display of high-resolution laser absorption spectra. Using a built-in computer(Linux OS)To provide continuous storage and measurement of data. With remote control function, users can operate the analyzer from any location through the network, and can also share real-time data and perform instrument diagnosis through remote login.
characteristic:
1. Real time synchronous measurementCO, CO2andH2Oconcentration
2. Higher accuracy
3. 10 HzHigh frequency measurement
4. Using a mid infrared quantum cascade laser, the measurement is not affected by changes in gas concentration or pressure
5. Full range linearity, with a range exceeding the environmental level20More than double
6. No need for liquid nitrogen or water cooling
Performance indicators:
Repeatability/Accuracy(1σ,1Seconds/100 seconds)
CO:0.1 ppb /0.3 ppb
CO2:0.2 ppm /0.08 ppm
H2O:3 ppm /1 ppm
Measurement speed:
10 Hz(Flow rate response>1 HzOptional external pump needs to be configured)
Drift(EPType,15Minute average, standard temperature and pressure,24Hour)
CO:0.5 ppb
CO2:0.1 ppm
H2O:10 ppmOr reading1%The larger one shall prevail
Measurement range (under all technical specifications):
CO:5~50000 ppb
CO2:2~10000 ppm
H2O:100 ppm~100% RH,non-condensing
Optional range:
CO:0~50 ppm
CO2:0~10000 ppm
H2O:0~100% RH,non-condensing
environment condition:
Sample temperature:0~50 ℃
Operating temperature:10~35 ℃(Standard type)/ 0~45 ℃(EPType)
ambient humidity:0~100% RH, without condensation
Temperature control accuracy(EPType):
0.003 ℃
Pressure control accuracy(EPType):
0.001 torr
Output:
Number(RS 232)Simulation, EthernetUSB
Electricity demand:
115/230 VAC,50/60 Hz,180 W(Standard type)/ 300 W(EPType)
Size and weight:
Standard type:22 cm(H)x 48 cm(W)x 80 cm(D),36 kg
EPType:49 cm(H)x 48 cm(W)x 80 cm(D),68 kg
Ordering information:
model(Model):
GLA251-CCQC(Standard type, rack mounted)
GLA351-CCQC(EPType, rack mounted)
Optional items:
1. MIU-16:16Road multiplexer
2. MIU-8:8Road multiplexer
3. 908-0008-9009:N920Vacuum pump (gas update time)1.2Seconds)
4. 908-0001-9011:N940Vacuum pump (gas update time)0.5Seconds)
5. 907-0005-9002Dynamic dilution system, capable of automatic dilution and range expansion100times
6. 904-0002Data collection software (includingUSB/RS 232Cable), capable of recording and synchronizing multiple devicesLGRAnalyzer or other equipment (such asGPSData output from anemometers, etc
manufacturer:
ABB Canada