Wuxi Dianchuang Technology provides you with instrument prices, models, user manuals, technical parameters, and more. Fully functional, stable performance, and easy to operate; The instrument adopts LCD display with background light, which has error indication and automatic temperature compensation.
Wuxi Dianchuang Technology Co., Ltd. DCT-EC0602 online conductivity meter has complete functions, stable performance, and easy operation; The instrument adopts LCD display with background light, which has error indication and automatic temperature compensation.
1Conductivity metercharacteristic
ØFully functional, stable performance, and easy to operate;
ØThe instrument adopts LCD display with background light and has error indication;
ØAutomatic temperature compensation;
ØIsolated 4-20 mA current output; Dual group relay control, adjustable hysteresis, high and low point alarm indication;
ØPower down memory, no need for backup batteries, data can be stored for more than ten years.
ØChoose an electrode constant of K=0.01 based on the range of conductivity of the tested water sample、0.1、1.0 or 10 electrodes shall be installed in a flow-through, submerged, flanged, or tubular manner.
IITECHNICAL INDEX
1.Measurement range:
Conductivity:0~19.99 uS/0~199.9 uS/0~1999 uS;0~19.99mS/0~199.9mS;
Electrical resistivity:0~19.99K/0~19.99M;
2.resolving power:0.01 uS 0.01M;
3.accuracy:0.02 uS;0.01M;
4.Stability: ≤0.04 uS/24h;≤0.02 M/24h;
5.Standard solution:Any standard solution;
6.Control range:0~19.99mS/0~19.99M/0~19.99M;
7.Temperature compensation:0~65.0℃;
8.Output signal: 4~20mAIsolation protection output;
9.Control output:ON/OFFRelay output contacts (two sets);
10.Relay bearing load:exchange230V5A;exchange115V10A;
11.Current output load:Allow load to be500;
12.Working voltage:230VAC10%、50/60Hz;
13.Size:96×96×110mmHole size: 92 * 92mm;
14.Environmental temperature:5~45℃.
Installation of conductivity meter:The instrument is suitable for embedded installation on the upper panel. The opening size is shown in Figure 1, which is 91 × 91mm. Insert the instrument into the square hole and fix it with the provided bracket.
Electrode installation:① Flow type (connected by hose or hard tube):

② Pipeline type: 
③ Flange type: Customized flange installation according to user needs.
④ Input type: Directly insert the electrode into the water tank, water tank, or water tower, waterproof the electrode cable or install it with a protective sleeve.
⑤ Other methods: If no suitable installation method is found above, please contact the manufacturer to customize a special installation method.
Online conductivity meter: http://www.wxdct.cn/c_html_products/1201-541.html
enclosure:《DCT-EC0602 Online Conductivity Meter User Manual 2016 Upgraded Edition》
Factors affecting conductivity measurement