Floor standing hot deformation Vicat tester with three workstations

NegotiableUpdate on 02/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Floor standing hot deformation Vicat tester with three workstations, using PLC+Windows system programmable controller control system, mainly used for measuring the hot deformation temperature and Vicat softening point temperature of non-metallic materials such as plastic, rubber, nylon, electrical insulation materials, etc. The product meets the requirements of IS075 (E), IS0306 (E), GB/T8802, GB/T1633, GB/T1634 and other standards. It is an instrument for self inspection by various quality inspection units, colleges and universities, and enterprises.

Product Details

Floor standing hot deformation Vicat tester (Vicat softening point temperature tester), floor standing hot deformation temperature tester, hot deformation Vicat tester, floor standing Vicat softening point temperature tester, floor standing hot deformation Vicat softening point temperature tester

purpose

1ApplicationPLC+WindowssystemProgrammable controller control system, mainly used for measuring the thermal deformation temperature and Vicat softening point temperature of non-metallic materials such as plastic, rubber, nylon, electrical insulation materials, etc. The product complies withIS075(E)TheIS0306(E)TheGB/T8802TheGB/T1633TheGB/T1634According to standard requirements, it is an instrument for self inspection by various quality inspection units, colleges and universities, and enterprises.

2The hot deformation temperature measuring instrument is easy to operate, convenient to use, has stable performance, high product accuracy, and can monitor the test temperature and deformation amount in real time during the test process; At the end of the experiment, the system automatically stops heating and can print and store the test report and test curve.

working principle

Plastic bending load thermal deformation temperature: When the temperature is uniformly increased, the temperature at which the upper surface of the thermoplastic specimen is bent and deformed to the specified deflection under the standard head pressure specified in the standard load conditions is measured.

execution standard

ISO75-11993Plastics - Determination of Load Deformation Temperature

ISO3061994Plastics - Determination of Vicat softening point temperature of thermoplastic materials

GB/T1633-2000Determination of Vicat softening point temperature of thermoplastic materials

GB/T1634-2004Plastics - Determination of Load Deformation Temperature

JB/T 12723thermal deformation/Technical specification for Vicat softening temperature tester

GB/T 1633Thermoplastic Vicat softening point temperature(VST)The determination

GB/T 1634.1Determination of plastic load deformation temperature1Part: General Experimental Methods

GB/T 1634.2Determination of plastic load deformation temperature2Part: Plastic and Hard Rubber

Q/SY 13474Management Specification for Quality Inspection of Goods Arrival

Q/SY 13281Standard for Material Storage Management

Technical Specifications

Control system:PLC+WindowsSystem upper computer

Operating interface: Color12Inch embedded industrial computer, switching between Chinese and English;

Testing stations: three stations and six stations (choose one from the two);

Workstation upgrade function: fully automatic lifting, safe, reliable, and highly stable;

printer:WiFiConnect printerA4

Data analysis function:Can display experimental data and generate temperature controlled displacement curve graphs

Temperature control range: oil temperature~300

Temperature resolution:0.1

Large temperature control error: ±1/6Minutes(120/h),

±0.5/6Minutes(50/h

Heating rate:120/h50/h50/h120/hSet arbitrarily between them.

Large deformation measurement range:1mm

Measurement error of large deformation: ±0.001mmrange0-12.7MM

Thermal deformation test span:64mm100mm

Hot deformation head and support radius:R3±0.02mm

Vicat needle area:1mm2

Vicat needle length:4mm

Heating medium: Methyl silicone oil (flash point)300Above ℃

High heating power:3.4KW

Control power supply (main power supply):220V±10%15A50Hz

Heating power supply (auxiliary power supply):220V±10%5A50Hz

Quality of Load Rod and Tray:76±2g

Equipment overall size:three workstations720*530*1100mm/six stations1020*530*1100mm

weight of the weight

Quantity per sample rack, unitg

2000g.500g.100g.50geach3blockOr each6According to workstation configuration

1000g.200g

Floor standing hot deformation Vicat tester with three workstations Floor standing hot deformation Vicat tester with three workstations

落地式热变形维卡测试仪三工位