G2205 gas concentration analyzer measures HF and H2O

NegotiableUpdate on 02/24
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Measure HF and H2O

Product Details

The Picarro G2205 gas concentration analyzer can accurately measure hydrogen fluoride (HF) and water vapor (H) in real-time2O) , The sensitivity is in parts per trillion (ppt), and the drift generated for atmospheric science and air quality applications can be ignored. This analyzer is equipped with coating components in key gas pathways, which reduces the tendency of HF molecules to adhere to the surface of the gas pathway, thereby improving the measurement response time. This analyzer requires almost no consumables, therefore it has significant advantages of ease of use and lower purchase cost. This analyzer has a sturdy structure and is easy to transport.

  • Fast, continuous, and real-time measurement of HF

  • Long term stability to meet low-frequency calibration requirements

  • Small footprint, suitable for on-site or laboratory deployment, no need for consumables

  • Equipped with water vapor correction function, automatically reporting dry gas mole fraction

The accuracy at 5 minutes is 30 ppt. The drift within 24 hours at standard temperature and pressure (STP) is less than ± 150 ppt. The moisture correction software will automatically report the dry gas mole fraction to help reduce the complexity of research work and save consumable costs.


Picarro G2205 Performance Specifications
Target gas
(In air like matrix)
Detection limit * (5 minutes, 3 σ) Zero drift * (within 24 hours/1 month)
(Peak to peak value of 5-minute mean)
HF 30 ppt ± 0.15 / ± 0.5 ppb

*After 1 hour of stable equilibrium, ensure the following operating conditions are met.

Picarro G2205 System Specification+A6: C27
measurement Cavity ring down spectroscopy (CRDS) technology
Measurement pool temperature control ± 0.005 ℃
Measurement pool pressure control ± 0.0002 atmospheric pressure
measuring range Ensure that the specification range is 0-0.5 ppm and the working range is 0-1.0 ppm
The accuracy for measuring higher concentrations is approximately 30 ppt+0.5% reading
Measurement Interval About 5 seconds
response time 90-10% decrease time, 0.3 ppm -0,<10='' minutes. 10-90%='' rise time, 0-0.3=''><10=''>
sample temperature -10 to 45 ℃
Sample flow rate Less than 0.6 standard liters per minute (slm) at 760 pallets, no filtration required
Sample pressure 300 to 1000 Torr (40 to 133 kPa)
Sample humidity Relative humidity (RH) is less than 99% (under non condensing conditions at 40 ℃). H2O can affect HF measurement. To achieve optimal performance, both drying treatment of HF and consistent H2O concentration can be provided
ambient temperature 10 to 35 ℃ (operating); -10 to 50 ℃ (storage)
ambient humidity Relative humidity (RH) less than 99%, under non condensing conditions
Other gases measured H2O、O2
attachment Pump (external, included), keyboard (included), mouse (included), LCD monitor (included)
data output RS-232、 Ethernet USB、 Simulation (optional) 0-10 volts
Air inlet connector ¼ inch Swagelok ® PFA connector
External dimensions Analyzer: 17 inches wide x 7 inches high x 17.5 inches long (43.2 x 17.9 x 44.6 centimeters), excluding 0.5 inch feet
External pump: 12.8 inches long x 8.9 inches high x 6.2 inches wide (32.4 x 22.6 x 15.8 centimeters)
Installation form Workbench or 19 inch rack mounted chassis
weight 70 pounds (31.75 kilograms), including pump
Power requirements 100-240 volts AC, 47-63 Hz (automatic detection), startup (total) less than 260 watts: 110 watts (analyzer),
At steady state, it is 120 watts (pump)
calibration The calibration of this instrument is limited by the preparation of constant concentration samples. We should investigate whether the instrument is suitable for commercial standards based on your application. For more details, please contact Picarro.