The Picarro G2301 gas concentration analyzer can accurately measure carbon dioxide (CO2), methane (CH4) with a sensitivity of parts per billion (ppb), and water vapor (H2O) with a sensitivity of parts per million (ppm) simultaneously. The drift generated for atmospheric science, air quality, and quantitative emission applications can be ignored. This analyzer meets the performance requirements of the World Meteorological Organization (WMO) and the Integrated Carbon Observing System (ICOS) for atmospheric monitoring of CO2 and CH4.
Simultaneous and continuous measurement of three atmospheric trace gases
Billion parts per billion (ppb) sensitivity, accuracy, and precision, with extremely low drift
Complies with the international environmental and atmospheric monitoring requirements of WMO and ICOS
Equipped with water vapor correction function, automatically reporting dry gas mole fraction
For CO2, the accuracies at 5 seconds and 5 minutes are<70 ppb='' and=''>70><25 ppb, respectively. For CH4, the accuracies are=''>25><0.5 ppb='' and=''>0.5><0.22 ppb, respectively. For='' CO2 and='' CH4, the maximum drift within 24 hours at standard temperature and pressure='' (STP)='' is='>0.22><120 ppb=' 'and='>120><1>, respectively
| Picarro G2301 ensures performance specifications (in air) | CO2 | CH4 | H2O | |
| Accuracy (1 σ) Ensure accuracy under the following specified operating conditions |
5 seconds | < 70=''> Typical value=15 ppb* |
< 0.5=''> Typical value=0.14 ppb* |
< 80=''> |
| 5 minutes | < 25=''> Typical value=5 ppb* |
< 0.22=''> Typical value=0.04 ppb* |
< 30=''> | |
| Maximum drift under standard temperature and pressure ***(The difference between the maximum values of the 50 minute average) Ensure drift under the following specified operating conditions |
24 hours | 120 ppb Typical value=17.0 ppb* |
1 ppb Typical value=0.10 ppb* |
<> |
| monthly | 500 ppb | 3 ppb | - | |
| Automatic determination of dry mole fraction | Already included | Already included | - | |
| measuring range | 0–1000 ppm | 0–20 ppm | 0-7% v (dew point at 39 ℃) Under non condensing conditions |
|
| Ensure accuracy range | 300–700 ppm | 1–3 ppm | 0-3% v (dew point at 25 ℃) Under non condensing conditions |
|
| Measurement interval (data acquisition rate) | < 5=''> Typical value=2.8 seconds* |
|||
| Gas response: rise/fall time (10-90%/90-10%) | < 3=''> | |||
**The typical performance is defined as the median of the test results of 36 G2301 analyzers manufactured in sequence. Results can be requested.
*Picarro calculates drift by subtracting the minimum value from the maximum value of the 50 minute average data obtained from a 30 hour experiment.
| Picarro G2301 ensures operating conditions | ||||
| measurement | Cavity ring down spectroscopy (CRDS) technology | |||
| Measurement pool temperature control | ±0.005 ℃ | |||
| Measurement pool pressure control | ± 0.0002 atmospheric pressure | |||
| sample temperature | -10 to 45 ℃ | |||
| Sample pressure | 300 to 1000 Torr (40 to 133 kPa) | |||
| Sample flow rate | <0.4 standard liters per minute='' (='' slm=''), at='' 760=''>0.4> | |||
| Sample humidity | <99% relative humidity (at='' 40='' ℃=''>99%) | |||
| ambient temperature | 10 to 35 ℃ (during instrument operation), -10 to 50 ℃ (instrument storage conditions) | |||
| ambient humidity | <85%>85%> | |||
| Accessories (included) | External vacuum pump, vacuum pipeline and accessories, keyboard, mouse, LCD display (optional), built-in 240GB hard drive | |||
| data output | RS-232, Ethernet, USB, Analog signal (optional) 0-10 volts | |||
| Air inlet connector | ¼ inch Swagelok ® | |||
| size | Analyzer: 17 inches wide x 7 inches high x 17.5 inches deep (43.2x17.8x44.5 centimeters), excluding 0.5 inch foot pads External pump: 7.5 inches wide x 4 inches high x 11 inches deep (19 x 10.2 x 28 centimeters) |
|||
| Installation form | Workbench (standard) or 19 inch rack mounted chassis (optional) | |||
| weight | 60.4 pounds (27.4 kilograms), including external pump | |||
| Power requirements | 100-240 volts AC, 47-63 Hz (automatic detection),<375>375> ~120 watts (analyzer)+150 watts (vacuum pump), during stable operation. |
|||