The Picarro G2308 gas concentration analyzer can accurately measure nitrous oxide (N2O), methane (CH4) with a sensitivity of parts per billion (ppb), and water vapor (H2O) with a sensitivity of parts per million (ppm) simultaneously. The drift generated for atmospheric science, air quality, and quantitative emission applications can be ignored. This analyzer can operate in open or closed systems and can be easily integrated with chamber systems.
Synchronous continuous measurement of N2O and CH4
Billion parts per billion (ppb) sensitivity, accuracy, and precision
Can operate in open or closed systems and easily integrate with chamber systems
Equipped with water vapor correction function, automatically reporting dry gas mole fraction
For N2O, the accuracies at 1 minute and 5 minutes are<10 ppb='' and=''>10><3.5 ppb, and for CH4=''>3.5><7 ppb='' and=''>7><3>3>
| Performance specifications of Picarro G2308 in air | |||
| specifications | N2O | CH4 | H2O |
| Initial accuracy (1 σ) | < 25='' ppb='' +='' 0.05%=''> | < 10='' ppb='' +='' 0.05%=''> | < 500=''> |
| Accuracy at 1 minute (1 σ) | < 10='' ppb='' +='' 0.05%=''> | < 7='' ppb='' +='' 0.05%=''> | < 250=''> |
| Accuracy at 5 minutes (1 σ) | < 3.5='' ppb='' +='' 0.008%=''> | < 3='' ppb='' +='' 0.02%=''> | < 100=''> |
| Ensure specification range | 0.3–200 ppm | 1–15 ppm | 0–3% |
| scope of work | 0–400 ppm | 0–20 ppm | 0–7% |
| measurement speed | < 6=''> | < 10=''> | < 8=''> |
| Typical gas response | < 10=''> | < 10=''> | - |
Analyzer specificity: Picarro cavity ring down spectroscopy (CRDS) technology uses an extremely narrow spectral region. Compared with other spectroscopic measurement techniques, this greatly reduces the possibility of interference from other gas components. But in actual samples, interference sometimes occurs. The Picarro analyzer is equipped with interference detection software and has been tested and characterized for its susceptibility to the following components:
| Picarro G2308 Trace Gas | Sensitivity of N2O to interference | ||
| carbon dioxide | No - Automatic calibration to 20000 ppm CO2 | ||
| methane | No - Automatic calibration to 200 ppm CH4 | ||
| ammonia | No - Automatic correction to 2 ppm NH3 | ||
| ethane | 0.2 ppb N2O / ppm C2H6 , Tested to 120 ppm | ||
| ethylene | 0.5 ppb N2O / ppm C2H4 , Tested to 16 ppm | ||
| acetylene | Not applicable for acetylene experiment | ||
| background gas | Designed for ambient gases, not suitable for highly variable or concentrated N2, O2, H2, or He background gases | ||
| ChemDetect ™ software | The unique Picarro algorithm can detect and label data that may be inaccurate due to spectral interference | ||
| Picarro G2308 analyzer operating parameters | specifications | ||
| ambient temperature | 10–35 ℃ | ||
| ambient humidity | Relative humidity (RH) less than 99%, under non condensing conditions | ||
| Sample pressure | 300 to 1000 Torr (40 to 133 kPa) | ||
| Sample flow rate | Approximately 230 standard milliliters per minute (sccm) | ||
| Sample humidity | Relative humidity (RH) less than 99%, under non condensing conditions, water vapor calibration test to dew point of 25 ℃ | ||
| sample temperature | -10–45 ℃ | ||
| Chamber temperature control | ± 0.005 ℃ | ||
| Chamber pressure control | ± 0.0002 atmospheric pressure | ||
| Closed circuit/circulation capability | Compatible with Picarro closed system pump A0702 | ||
| Air inlet connector | ¼ inch Swagelok ® | ||
| External dimensions | 17 inches wide x 7 inches high x 17.5 inches long (43.2 x 17.8 x 44.6 centimeters), excluding 0.5 inch feet | ||
| weight | 50 pounds (22.6 kilograms) | ||
| Power requirements | 100-240 volts AC, 47-63 Hz (automatic detection), less than 260 W at startup; 110 W at steady state | ||
| Installation form | Workbench (standard) or 19 inch rack mounted chassis (optional) | ||
| attachment | Accompanying: keyboard, mouse. Optional: LCD monitor. Excluding: Vacuum pump | ||
| option | A0702, Picarro Closed System Pump S0528, O2 sensor, used for O2 measurement and calibration in constantly changing O2 environments S0517, CH4 extended working range up to 800 ppm |
||