G2308 gas concentration analyzer measures N2O, CH4, and H2O

NegotiableUpdate on 02/24
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
G2308 gas concentration analyzer measures N2O, CH4, and H2O
Product Details

The Picarro G2308 gas concentration analyzer can accurately measure nitrous oxide (N2O), methane (CH4) with a sensitivity of parts per billion (ppb), and water vapor (H2O) with a sensitivity of parts per million (ppm) simultaneously. The drift generated for atmospheric science, air quality, and quantitative emission applications can be ignored. This analyzer can operate in open or closed systems and can be easily integrated with chamber systems.


  • Synchronous continuous measurement of N2O and CH4

  • Billion parts per billion (ppb) sensitivity, accuracy, and precision

  • Can operate in open or closed systems and easily integrate with chamber systems

  • Equipped with water vapor correction function, automatically reporting dry gas mole fraction



For N2O, the accuracies at 1 minute and 5 minutes are<10 ppb='' and=''>10><3.5 ppb, and for CH4=''>3.5><7 ppb='' and=''>7><3>3>



Performance specifications of Picarro G2308 in air
specifications N2O CH4 H2O
Initial accuracy (1 σ) < 25='' ppb='' +='' 0.05%=''> < 10='' ppb='' +='' 0.05%=''> < 500=''>
Accuracy at 1 minute (1 σ) < 10='' ppb='' +='' 0.05%=''> < 7='' ppb='' +='' 0.05%=''> < 250=''>
Accuracy at 5 minutes (1 σ) < 3.5='' ppb='' +='' 0.008%=''> < 3='' ppb='' +='' 0.02%=''> < 100=''>
Ensure specification range 0.3–200 ppm 1–15 ppm 0–3%
scope of work 0–400 ppm 0–20 ppm 0–7%
measurement speed < 6=''> < 10=''> < 8=''>
Typical gas response < 10=''> < 10=''> -

Analyzer specificity: Picarro cavity ring down spectroscopy (CRDS) technology uses an extremely narrow spectral region. Compared with other spectroscopic measurement techniques, this greatly reduces the possibility of interference from other gas components. But in actual samples, interference sometimes occurs. The Picarro analyzer is equipped with interference detection software and has been tested and characterized for its susceptibility to the following components:

Picarro G2308 Trace Gas Sensitivity of N2O to interference
carbon dioxide No - Automatic calibration to 20000 ppm CO2
methane No - Automatic calibration to 200 ppm CH4
ammonia No - Automatic correction to 2 ppm NH3
ethane 0.2 ppb N2O / ppm C2H6 , Tested to 120 ppm
ethylene 0.5 ppb N2O / ppm C2H4 , Tested to 16 ppm
acetylene Not applicable for acetylene experiment
background gas Designed for ambient gases, not suitable for highly variable or concentrated N2, O2, H2, or He background gases
ChemDetect ™ software The unique Picarro algorithm can detect and label data that may be inaccurate due to spectral interference


Picarro G2308 analyzer operating parameters specifications
ambient temperature 10–35 ℃
ambient humidity Relative humidity (RH) less than 99%, under non condensing conditions
Sample pressure 300 to 1000 Torr (40 to 133 kPa)
Sample flow rate Approximately 230 standard milliliters per minute (sccm)
Sample humidity Relative humidity (RH) less than 99%, under non condensing conditions, water vapor calibration test to dew point of 25 ℃
sample temperature -10–45 ℃
Chamber temperature control ± 0.005 ℃
Chamber pressure control ± 0.0002 atmospheric pressure
Closed circuit/circulation capability Compatible with Picarro closed system pump A0702
Air inlet connector ¼ inch Swagelok ®
External dimensions 17 inches wide x 7 inches high x 17.5 inches long (43.2 x 17.8 x 44.6 centimeters), excluding 0.5 inch feet
weight 50 pounds (22.6 kilograms)
Power requirements 100-240 volts AC, 47-63 Hz (automatic detection), less than 260 W at startup; 110 W at steady state
Installation form Workbench (standard) or 19 inch rack mounted chassis (optional)
attachment Accompanying: keyboard, mouse. Optional: LCD monitor. Excluding: Vacuum pump
option A0702, Picarro Closed System Pump
S0528, O2 sensor, used for O2 measurement and calibration in constantly changing O2 environments
S0517, CH4 extended working range up to 800 ppm