The Picarro G2401-m gas concentration analyzer can synchronously and accurately measure carbon monoxide (CO), carbon dioxide (CO2), and methane (CH) with a sensitivity of parts per billion (ppb)4)Water vapor (H) with a sensitivity of parts per million (ppm)2O) , The drift generated by atmospheric science, air quality, and quantitative emission applications can be ignored. Flight optimization design components minimize the impact of aircraft vibration, pitch, roll, and rapidly changing environmental conditions. Therefore, this analyzer can be operated on aircraft to achieve urban surveying or vertical profile analysis.
Simultaneous and continuous measurement of four atmospheric trace gases
Billion parts per billion (ppb) sensitivity, accuracy, and precision, with extremely low drift
Robust design ensures performance during flight
Equipped with water vapor correction function, automatically reporting dry gas mole fraction
For CO, CO2And CH4The accuracy at 5 minutes under vibration conditions of 20 Hz, 1 g, and 60 minutes is ≤ 30 ppb, ≤ 200 ppb, and ≤ 2 ppb, respectively. For CO, CO2And CH4The drift within 24 hours at standard temperature and pressure (STP) is ≤ 15 ppb, ≤ 200 ppb, and ≤ 1.5 ppb, respectively. With the continuous changes in temperature and pressure (see data manual for details), the drift remains relatively low. The moisture correction software will automatically report the dry gas mole fraction to help reduce the complexity of research work and save consumable costs.
| Picarro G2401-m analyzer specifications | CO2 specifications | CH4 specification | CO specification | H2O specification |
| Original accuracy (1- σ, 300 seconds, 20 Hz vibration, 1 g): Ensure accuracy within the following range and operating conditions | ≤ 200 ppb | ≤ 2 ppb | ≤ 30 ppb | ≤ 150 ppm |
| Drift at standard temperature and pressure (within 24 hours) (The worst value of the 50 minute mean: Ensure drift within the following range and operating conditions) |
≤ 200 ppb | ≤ 1.5 ppb | ≤ 15 ppb | ≤ 100 ppm+5% reading |
| Drift during continuous temperature changes (the difference between the 30 second average within 3 hours; the temperature change rate is 15 ℃ per hour under the following operating conditions) | ≤ 7.5 ppbv / ℃ | ≤ 0.05 ppbv / ℃ | ≤ 1.5 ppb / ℃ | Not Applicable |
| Drift during constant pressure changes (The worst value difference of the 30 second average; the pressure change rate is less than 1.4 Torr per second under the following operating conditions) |
≤ 700 ppb | ≤ 7.5 ppb | ≤ 50 ppb | Not Applicable |
| scope of work | 0–1000 ppm | 0–20 ppm | 0–5 ppm | 0–7 %v H2O / 39 ℃ dew point (no condensation) |
| Ensure specification range | 300–500 ppm | 1 ppm–3 ppm | 0–1 ppm | 0–3 %v H2O |
| Measurement Interval | ≤ 3.5 seconds | ≤ 3.5 seconds | ≤ 3.5 seconds | ≤ 3.5 seconds |
| Rise time/fall time (10-90%/90-10%) | ≤ 3 seconds | ≤ 3 seconds | ≤ 3 seconds | Not Applicable |
| Picarro G2401-m analyzer specifications | ||||
| measurement | Cavity ring down spectroscopy (CRDS) technology | |||
| Measurement pool temperature control | ±0.005 ℃ | |||
| Measurement pool pressure control | ± 0.0002 atmospheric pressure | |||
| sample temperature | -10 ℃ to 45 ℃ | |||
| Sample flow rate | Less than 0.6 standard liters per minute (slm) when exceeding 250-1000 pallets, no filtration required | |||
| Sample pressure | 250 to 1000 Torr (40 to 133 kPa) | |||
| Sample humidity | Relative humidity (RH) less than 99%, under non condensing conditions at 40 ℃ | |||
| ambient temperature | Operating temperature ranges from 10 ℃ to 35 ℃ (storage temperature ranges from -10 ℃ to 50 ℃) | |||
| Maximum rate of change in ambient temperature | 15 ℃ per hour | |||
| ambient humidity | Relative humidity (RH) less than 99%, under non condensing conditions | |||
| maximum height | Height at 250 pallets | |||
| Maximum rate of change in height | 1000 meters per minute | |||
| Attachment (attached) | Pump (external), keyboard, mouse, LCD monitor (optional) | |||
| output | RS-232、 Ethernet USB、 Simulation (optional) 0-10 volts | |||
| Air inlet connector | ¼ inch Swagelok ® | |||
| External dimensions | Analyzer: 17.6 inches long x 7 inches high x 17 inches wide (44.6 x 17.8 x 43.2 centimeters), excluding 0.5 inch feet External pump: 12.8 inches long x 8.9 inches high x 6.2 inches wide (32.4 x 22.6 x 15.8 centimeters) |
|||
| weight | 70 pounds (31.75 kilograms), including pump | |||
| Power requirements | 100-240 volts AC, 47-63 Hz (automatic detection). At startup (total) less than 260 watts: 110 watts (analyzer), steady state is 120 watts (pump) | |||
| Steady state power loss | < 370=''> | |||