Phone
13910499766
-
Address
Building 5, Guanghua Entrepreneurship Park, No. 18 Anningzhuang East Road, Haidian District, Beijing
Beijing Lijia United Technology Co., Ltd
13910499766
Building 5, Guanghua Entrepreneurship Park, No. 18 Anningzhuang East Road, Haidian District, Beijing
The Picarro G5310 gas concentration analyzer can accurately measure nitrous oxide (N2O), carbon monoxide (CO) with a sensitivity of parts per trillion (ppt), and water vapor (H2O) with a sensitivity of parts per million (ppm) simultaneously. The drift generated by atmospheric science, air quality, and quantitative emission applications can be ignored. This analyzer meets the performance requirements of the World Meteorological Organization (WMO) and the Integrated Carbon Observing System (ICOS) for atmospheric monitoring of N2O and CO.
Synchronous continuous measurement of N2O and CO
Mid infrared cavity ring down spectroscopy (CRDS) for high-precision and low drift analysis
Complies with the international environmental and atmospheric monitoring requirements of WMO and ICOS
Equipped with water vapor correction function, automatically reporting dry gas mole fraction

For N2O and CO, the accuracy at 5 seconds and 5 minutes is<200 ppt='' and=''>200><40 ppt, respectively. For='' n2o='' and='' co, the drift within 24 hours at standard temperature and pressure='' (stp)='' is less than='' 100=''>40>
| Picarro G5310 Performance Specifications | N2O | CO | H2O |
| Accuracy (1)σ, 5 seconds) | < 0.2='' (='' n2o=''>< 500='' ppb=''> Typical value=0.10 ppb* |
< 0.2='' (='' co=''>< 400='' ppb=''> Typical value=0.13 ppb* |
< 40=''> |
| Accuracy (1)σ, 5 minutes) | < 0.04='' ppb='' (='' n2o=''>< 500='' ppb=''> Typical value=0.015 ppb* |
< 0.04='' ppb='' (='' co=''>< 400='' ppb=''> Typical value=0.019 ppb* |
< 6=''> |
| Accuracy (1)σ, 1 hour) | < 0.02='' ppb='' (='' n2o=''>< 500='' ppb=''> | < 0.02='' ppb='' (='' co=''>< 400='' ppb=''> | - |
| Accuracy (1)σ, 5 minutes) | 0.03% N2O reading (N2O>500 ppb) | 0.02% CO reading (CO>400 ppb) | < 6=''> |
| Drift (24 hours) | < 0.1=''> Typical value=0.036 ppb* |
< 0.1=''> Typical value=0.046 ppb* |
- |
| measuring range | 1–1500 ppb | 1–1500 ppb | 0–3% |
| Measurement Interval | < 5=''> Typical value=2.7 seconds* |
||
*The typical performance is defined as the median of the test results of 41 G5310 analyzers manufactured in sequence. Results can be requested.
| Picarro G5310 System Specifications | |||
| measurement | Cavity ring down spectroscopy (CRDS) technology | ||
| Measurement pool temperature control | ±0.005 ℃ | ||
| Measurement pool pressure control | ± 0.0002 atmospheric pressure | ||
| sample temperature | -10 to 45 ℃ | ||
| Sample pressure | 300 to 1000 Torr (40 to 133 kPa) | ||
| Sample flow rate | 100 standard milliliters per minute | ||
| Sample humidity | <99% relative humidity (at='' 40='' ℃=''>99%) | ||
| ambient temperature | 15 to 35 ℃ (working conditions) -10 to 50 ℃ (storage conditions) |
||
| ambient humidity | < 85%=''> | ||
| Accessories (included) | Keyboard, mouse, LCD display (optional) | ||
| data output | RS-232 interface, network interface, USB interface | ||
| Air inlet connector | ¼ inch Swagelok ® | ||
| size | 17 inches wide x 12.5 inches high x 20.5 inches long (43.2 x 31.7 x 52.1 centimeters, excluding 0.5 inch foot pads) | ||
| Installation form | workbenches | ||
| weight | 95 pounds (43 kilograms), including built-in pump | ||
| Power requirements | 100-240 volts AC, 47-63 hertz (auto detect), 375 watts at startup and 200 watts at runtime | ||
Compatible peripheral devices: 16 channel gas injection system (A0311)
Note on deployability: G5310 analyzer is only designed for static installation and is not suitable for mobile platforms.
If field deployment is required, please consult Picarro for the configuration of the DC power supply.