This instrument adopts advanced microfluidic infrared gas sensor technology and can simultaneously measure the concentrations of five gases including SO2, NO, CO, CO2, O2, etc. It can also measure smoke parameters such as smoke flow rate and temperature.
Functional Features
◆Gaseous moisture in flue gasSO2、NOMeasurement has no impact
◆Smoke sampling flow rateSO2、NO、CO、CO2no effect
◆Built in imported air pump sampling
◆The sampling pipeline adopts imported heating and constant temperature pipeline
◆Automatic calculationLambdaAnd combustion efficiency
◆Configuration(320×240)LCD display screen, standardRS232Serial communication interface
Technical Parameter
◆ measuring range:
SO2:0-2000ppm , NO:0-2000ppm ,
CO: 0-5000ppm , CO2:0-25% , O2:0-25%,
Smoke velocity:0-40m/s ,flue gas temperature:0-600℃
◆SO2、NO、CO、CO2repeatability:±1%FS,O2repeatability: ±1%FS
SO2、NO、CO、CO2Indication error:±1%FS , O2Indication error:±2%FS
SO2、NO、CO、CO2Zero point and range drift:±1%FS ,
O2Zero point and range drift:±2%FS
◆ flue gas temperature: 0-600℃ Temperature error:±2℃Resolution:1℃
◆ Smoke velocity: 0-40m/sFlow velocity error:±2.5%FSResolution:0.1 m/s