JKZC-GType 6Non contact electrostatic voltageoutsideCalibration device
Keywords: safe operation box, non-contact, electrostatic voltmeter calibration
JKZC-GType 6Non contact electrostatic voltageoutsideCalibration deviceAdopting a safety shielding box to avoid contact between people and high voltage, convenient, safe and fast,Can quickly and accurately measure, calibrate, and verify most of the current domestic and foreign electrostatic voltmeters and electrostatic field strength meters. The standard type is mainly suitable for industries related to electronic components and electronic equipment such as electronics, aerospace, weapons, and national defense. High voltage non-contact electrostatic voltmeters are mainly suitable for calibration of electrostatic voltmeters in petroleum, chemical, oil depot, fire protection, natural gas, printing, textile, dyeing, rubber, plastic, production, storage and transportation safety management departments.
2、 Complies with:
*Complies with the national military standard GJB "Calibration Regulations for Non Contact Static Voltage Meters";
Compliant with standard: JJF 1517-2015 non-contact electrostatic voltage measuring instrument
3、 Product features:
1. It can calibrate both contact type electrostatic voltmeters and non-contact electrostatic voltmeters for measurement;
2. * Complies with the national military standard GJB "Calibration Regulations for Non contact Static Voltage Meters";
3. It can calibrate both the electrostatic voltage and the electrostatic field strength;
4. Digital display, high resolution, resolutionReaching 1V andHigh accuracy;
5. Good stability, small size, and light weight;
6. Continuously adjustable from 0, with a wide measurement range;
7. Multiple protection, safe and reliable operation;The scale error is less than ± (0.01-0.1) mm, and the display unit can be in inches (convenient for measuring electrostatic meters with inch specified distances in Europe and America), or it can be switched to directly display mm (convenient for measuring electrostatic meters with unit specified distances in China)
8. Multi functional fixture, capable of clamping probes for circular, square, and flat electrostatic voltmeters and field strength meters;
9. Complete set including electrostatic standard source, standard electrode, positioning bracket, etc
10. If the scale error is less than ± (0.01-0.1) mm, the display unit can be inches or switched to directly display mm
11. Provide metrology certificates from the China Institute of Metrology to ensure product performance.
12. Can calibrate positive and negative polarity electrostatic voltmeter calibration device
13. The entire equipment is operated inside a secure shielding box to ensure the personal safety of personnel
JKZC-G3Specification
measuring rangeadjustable 0~±20Kv, 0~±30Kv, 0~±60Kv, multiple options to choose from
Range: 0.05level
Accuracy: 0.01%
Positive high voltage power supply
Output voltage range 0~±20Kv
b: Negative high voltage power supply
Output voltage range 0~±20Kv
Small resolution ±1V
Stability 0.01%
Voltage regulation fineness ≤1V
Scale range (1~300)mm;
Voltage output error Better than ± 1%
Scale error is better than ±(0.01~1)mm
The insulation resistance of the insulation bracket is greater than 100TΩ
The withstand voltage of the insulation bracket>50kV
The adjustment fineness of the distance adjustment device is ≤ 001mm
Safe operation box: fully enclosed safety protection device size: 900 * 600 * 850