Maintenance method for dental consistency and strain testing device

NegotiableUpdate on 02/18
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The maintenance method of the dental consistency and compression strain testing device meets the requirements of YY/T 0493-2022 "Dental Elastic Impression Materials" for consistency and compression strain testing

Product Details

Maintenance method for dental consistency and strain testing device

Maintenance method for dental consistency and strain testing device


Dental viscosity and strain testing device

1、 Overview:

This device meets the requirementsYY/T 0493-2022Requirements for Consistency and Compressive Strain Testing of Dental Elastic Impression Materials

2、 Technical parameters:

parameter

Specifications

Consistency test load

(14.71±0.01)N

Working time load of "0" type material

(12.75±0.01)N

Compressive strain load

(12.26±0.01)N

Load rod

Provide an initial force of (1.22 ± 0.01N) for compressive strain

counterweightcontrol accuracy

±0.01N

Range of dial gauge

0~5mm

Precision of dial gauge

0.01mm

Operating Temperature

0~30℃

Operating Humidity

≤60%RH



3、 Execution standards

YY/T 0493-2022Dental elastomer impression materials

4、 Device principle


5、 Configuration List:

1 host;

1 copy of instruction manual;

1 certificate of conformity;

1 warranty card;

One copy of the delivery receipt;

1 nameplate;

1 set of wrench;


Dental viscosity and strain testing device

1、 Overview:

This device meets the requirementsYY/T 0493-2022Requirements for Consistency and Compressive Strain Testing of Dental Elastic Impression Materials

2、 Technical parameters:

parameter

Specifications

Consistency test load

(14.71±0.01)N

Working time load of "0" type material

(12.75±0.01)N

Compressive strain load

(12.26±0.01)N

Load rod

Provide an initial force of (1.22 ± 0.01N) for compressive strain

counterweightcontrol accuracy

±0.01N

Range of dial gauge

0~5mm

Precision of dial gauge

0.01mm

Operating Temperature

0~30℃

Operating Humidity

≤60%RH



3、 Execution standards

YY/T 0493-2022Dental elastomer impression materials

4、 Device principle


5、 Configuration List:

1 host;

1 copy of instruction manual;

1 certificate of conformity;

1 warranty card;

One copy of the delivery receipt;

1 nameplate;