In 2011,LGR(Currently under ABB Canada)Released its new version of nitrous oxide/carbon monoxide analyzer (N2O/CO-23r)Can simultaneously measure N in the ambient atmosphere2O. CO and H2O concentration, and record the dry and wet mole fractions without the need for sample drying. The analyzer is designed for various purposes, including trace gas monitoring, eddy related flux measurement, box method flux measurement, exhaust emission measurement, etc. The analyzer is particularly suitable for field measurements and has been successfully installed on NASA's DC-8 spacecraft for measuring the atmosphere in the troposphere and stratosphere. N2O/CO-23rAdopting built-in temperature control technology, it provides higher precision and accuracy, as well as smaller drift.
N2O/CO-23rThe off-axis integrated cavity output spectroscopy (OA-ICOS) technology designed with LGR patent (patent number: 6795190) eliminates the disadvantage of CRDS technology that requires continuous matching of optical cavity and laser wavelength during measurement to improve weak signal intensity. This eliminates the need for complex laser alignment adjustment, temperature control, and wavelength monitoring in the analyzer. Real time display of high-resolution laser absorption spectra. Adopting a built-in computer (Linux OS) to provide continuous storage and measurement of data. With remote control function, users can operate the analyzer from any location through the network, and can also share real-time data and perform instrument diagnosis through remote login.
characteristic:
1. NOAA Standard Designated Equipment for Atmospheric Background Monitoring and Vortex Related Flux Measurement
2. Real time synchronous measurement of N2O. CO and H2O concentration
3.moreHigh precision: The measurement accuracy can reach 0.1 ppb per second
4. 10 Hz high-frequency measurement
5. Range exceeding the environmental level by more than 20 times, with full range linearity
6. Adopting mid infrared quantum cascade laser, unaffected by other gases or pressure changes
7. No need for liquid nitrogen or water cooling
Performance indicators:
repeatability/Accuracy(1σ, 0.1 sec / 1sec / 100sec)
N2O:0.3 ppb/0.08 ppb/0.04 ppb (standard type, EP type)
CO:0.3 ppb / 0.1 ppb / 0.050 ppb(Standard type,EPType)
H2O:150 ppm / 50 ppm / 10 ppm(Standard type,EPType)
The accuracy of portable instruments is standard2 times
The data rate is less than1 Need fast streaming option in seconds
Measurement speed:
0.01~10 Hz (flow rate response>1 Hz requires optional external pump configuration)
Drift(EPType,15Minute average, standard temperature and pressure,24Hour)
N2O:1 ppb
CO:1 ppb
H2O: 50 ppm or 1% of the reading, whichever is greater
Measurement range (under all technical specifications):
N2O:1~4000 ppb
CO:1~4000 ppb
Optional range:
N2O:0~10 ppm
CO:0~10 ppm
Accuracy:
Uncertainty<1%, without calibration conditions (standard type)
Uncertainty<0.03%, without calibration conditions (EP type)
Environmental conditions:
Sample temperature: -30~50 ℃
Operating temperature: 10~35 ℃ (standard type)/0~45 ℃ (EP type)
Environmental humidity: 0~98% RH, no condensation
Output:
Digital (RS 232), Analog, Ethernet USB
Electricity demand:
115/230 VAC, 50/60 Hz, 180 W (standard type)/300 W (EP type)
Size and weight:
Standard Type: 35.6 cm (H) x 48.3 cm (W) x 76.2 cm (D), 36 kg
EP type: 48.9 cm (H) x 48.3 cm (W) x 86.4 cm (D), 68 kg
Portable type:20.3 cm(H)x 50.8 cm(W)x 61 cm(D),23 kg
performance testing:
N2O/CO-23rReal time provides accurateN2OandCOMeasurement function(Dry and wet values),No need for any correction or post analysis
Ordering information:
Model number(Model):
GLA251-N2OCM (Standard Type)
GLA351-N2OCM(EP type)
GLA151-N2OCM(Portable)
Optional items:
1. MIU-16:16 channel multiplexer
2. MIU-8: 8-channel multiplexer
3. 908-0008-9009: N920 vacuum pump (gas update time 1.2 seconds)
4. 908-0001-9011: N940 vacuum pump (gas update time 0.5 seconds)
5. 907-0005-9002: Dynamic dilution system, capable of automatic dilution and expanding the range by 100 times
6. 904-0002: Data acquisition software (including USB/RS 232 cable), capable of recording and synchronizing data output from multiple LGR analyzers or other devices (such as GPS, anemometer, etc.)
manufacturer:
ABB Canada
Application Literature
'Impact of extreme precipitation and water table change on N2O fluxes in a bio-energy poplar plantation'
D. Zona, I. A. Janssens, M. S. Verlinden, L. S. Broeckx, J. Cools, B. Gioli, A. Zaldei, and R. Ceulemans
Biogeosciences Discuss, volume 8, 2057–2092, 2011
'Development of a Fast, High-Precision Analyzer for N2O and CO Measurements in Field Applications'
Robert Provençal, Yonggang He, Raymond Fellers, Thomas Owano, Doug Baer
European Geophysical Union General Assembly, May 3 - 7, 2010
'Cavity-enhanced quantum-cascade laser-based instrument for carbon monoxide measurements'
R. Provencal, M. Gupta, T. G. Owano, D. S. Baer, K. N. Ricci, A. O'Keefe, and J. R. Podolske
Applied Optics, Vol. 44, 6712-6717 (2005)
'High-Frequency Isotope Measurements in N2O by Using Mid-IR Laser Absorption Spectroscopy'
Dong, F.; Baer, D. S.
American Geophysical Union, Fall Meeting 2010, abstract #B23A-0384