Picarro PI2114 can measure hydrogen peroxide (H2O2) as low as 3 ppb to prevent drug oxidation and ensure drug stability. Major pharmaceutical, contract manufacturing outsourcing (CMO), and isolator companies use Picarro hydrogen peroxide analyzers to achieve efficient API and biopharmaceutical production, as well as sterile filling and processing. The PI2114 analyzer runs quickly and is easy to use, without the need for chemicals or consumables. At the same time, only a small amount of calibration and maintenance is required to minimize operating costs.
Continuous real-time monitoring of H2O2
Detection limit:<3 ppb (5=''>3>)
Accuracy:<0.5 ppb (1 σ, 5=''>0.5>)
The rise time (10-90%) and fall time (90-10%) are within 2 minutes
The precision of 5 minutes is less than 0.5 ppb, the rise time (10-90%) and fall time (90-10%) are less than 2 minutes, and the detection limit is less than 3 ppb (5-minute average). The Picarro PI2114 hydrogen peroxide concentration analyzer has been optimized for GMP applications and its software complies with electronic record regulations. The built-in software of PI2114 complies with the requirements of 21 CFR part 11, and the software has strict regulations on electronic records, time, and audit tracking. The PI2114 software system, combined with detailed validation reports, software, and system files, provides a comprehensive and strictly controlled data management function that meets pharmaceutical and commercial manufacturing requirements. Picarro provides installation and operation certification documents (IQ and OQ), and features fast and easy validation using commercially available proxy gases.
| Picarro PI2114 Performance Specifications | Technical specifications** | ||||
| Accuracy (1 σ, 10 seconds) Accuracy (1 σ, 300 seconds) |
<3='' ppb=''+'' reading=''> <1 ppb=''+='' reading=''>1> |
||||
| Detection limit (3 σ, 300 seconds) | < 3.0=''> | ||||
| Zero drift * (72 hours) Zero drift (1 month) |
± 2.5 ppb - 5 ppb/+10 ppb |
||||
| accuracy | ± 5% of the reading | ||||
| measuring range | 0–100 ppm | ||||
| Measurement Interval | About 10 seconds | ||||
| response time (Rise time/Fall time 10-90%/90-10%) |
< 1=''> | ||||
*The Picarro analyzer does not require the use of zero standard gas or zeroing modules to operate or meet specification requirements.
**Each ordered analyzer will provide technical specifications and instrument specific test reports (compliance certificates).
| Picarro PI2114 System Specifications | |||||
| measurement | cavity ring-down spectroscopy | ||||
| Measurement pool temperature control | ±0.005°C | ||||
| Measurement pool pressure control | ± 0.0002 atmospheric pressure | ||||
| sample temperature | -10 to 45 ° C | ||||
| Sample flow rate | 760 pallets less than 1 standard liter per minute (slm), no filtration required | ||||
| Sample pressure | 300 to 1000 pallets (40 to 133 kPa) | ||||
| Sample humidity | Relative humidity (RH) less than 99%, no condensation at 40 ° C, no need to dry | ||||
| ambient temperature | 10 to 35 ° C (operation) -10 to 50 ° C (storage) | ||||
| ambient humidity | Relative humidity (RH) less than 99%, non condensing | ||||
| Other gases measured | H2O、CH4 | ||||
| attachment | Pump (external, included), keyboard and mouse (included), LCD monitor (optional) | ||||
| communication interface | RS-232、 Ethernet USB、 Simulate 4-20 mA | ||||
| Pipe joint | ¼”Swagelok ® PFA connector | ||||
| External dimensions | Including the legs, it measures 17 inches wide by 8.38 inches high by 21 inches long (43.2 by 17.9 by 53.3 centimeters), 7.5 inches wide x 4 inches high x 11 inches long (19 x 10.2 x 28 centimeters) when excluding small external pump modules |
||||
| weight | 73 pounds (33.2 kilograms), including pump | ||||
| Power requirements | 100-240 volts AC, 47-63 hertz (automatic detection), and less than 260 watts at startup (total). In steady state, it is 110 watts (analyzer), 35 watts in steady state (pump) |
||||
| authentication | CE Mark、CDRH | ||||
| Country of manufacture | America | ||||
| System Verification | Including software wizards for analyzer validation, using CH4 as the standard proxy gas | ||||
Software and User Account Management
| function | Not logged in | login | operator | technician | administrators |
| Configure menu/settings (Additional login is required to access authentication or user management) |
✔ | ||||
| View Data Viewer | ✔ | ✔ | ✔ | ✔ | |
| Change display data | ✔ | ✔ | ✔ | ✔ | |
| Exit measurement | ✔ | ✔ | ✔ | ||
| Configure 4-20mA output | ✔ | ||||
| H2O2 verification | ✔ | ✔ | |||
| Change display data (Full access permission) |
✔ | ||||
| User Management | ✔ |