- Phone
-
Address
2nd Floor East, Building A2, Shunheda Industrial Park, Liuxiandong Industrial Zone, Xili Street, Nanshan District, Shenzhen
Shenzhen Changhong Technology Co., Ltd
2nd Floor East, Building A2, Shunheda Industrial Park, Liuxiandong Industrial Zone, Xili Street, Nanshan District, Shenzhen

This product utilizes sealed catalytic digestion followed by colorimetric measurement. The instrument adopts advanced cold light source, narrowband interference technology, and microcomputer automatic data processing to directly display the COD (mg/L) value of the sample. This instrument is widely used in environmental monitoring, sewage treatment, colleges, research institutions and other departments
|
【 Host 】 | |||
| Measurement range: | 5-2000mg/L (exceeding dilution determination) | Indication error: | ≤±5 % |
| Repeatability: | ≤3% | Anti chlorine interference: | ≤2000mg/L |
| Temperature control accuracy: | ±0.5℃ | Temperature control time: | 1~9999min adjustable |
| Temperature control system: | The room temperature can be set between 200 ℃, and the digestion temperature is 165 ℃. | Optical stability: | The absorbance value of the instrument drifts less than 0.002A within 20 minutes |
| Dissolving time: | 15min | Batch processing capacity: | 16 water samples |
| External dimensions: | Host: 340mm × 250mm × 130mm; Dissolver (DIS-16B type) 355mm × 260mm × 135mm | ||
| Weight: | Host 4kg Dissolver 5.5kg | Power consumption: | <500W |
| Normal usage conditions: | |||
| Environmental temperature: | 5~40℃ | Relative humidity: | ≤85% |
| Power supply: | AC(220±22)V;(50±0.5)Hz | No significant vibration or electromagnetic interference, avoid direct sunlight. | |
|
The measuring instrument is a cold light source and narrowband interferometric optical system with good optical stability. The digestion instrument adopts a 3.5-inch color touch screen with a resolution of 320x240 and displays in both Chinese and English. Simple and time-saving operation. Dissolving colorimetric does not require changing the tube. With multi-point error free calibration, zero error measurement of the set point is achieved, completely eliminating the errors caused by the nonlinearity of temperature sensors in common digesters on the market. The digestion instrument has a timed reminder and can be set to automatically shut down after digestion. The instrument has a secondary protection function. When the temperature of the digestion instrument exceeds 200 degrees, the instrument will cut off the power by itself to ensure the safety of the digestion process. |
![]() It can save 100 standard curves and 10000 measured values each, without losing them when powered off. LCD large screen LCD display, easy and intuitive operation. |
![]() Printing function: It can print test records immediately or query records for printing. USB interface, can be connected to a computer. |
| name | quantity | name | quantity |
| host | 1 unit | test tube rack | 1 piece |
| Digester | 1 unit | COD reagent | 1 set |
| Dissolving colorimetric tube | 10 pieces | Dissolve the protective cover | 1 piece |
| usb cable | 1 piece | Data collection installation CD | 1 piece |
| power line | 2 pieces | operation instructions | 1 copy |
| Product Qualification Certificate | 1 copy | warranty card | 1 copy |