Ammonia gas(NH3)Gas concentration analyzer
●Fast, continuous, and real-time measurement
●Ultra high sensitivity, precision, and accuracy
● Good long-term stability, no need for frequent calibration
● Measure water(H2O)And carbon dioxide(CO2)To calibrate and validate
● Small size, can be deployed in the field or laboratory, no consumables required
PicarroSI2103The gas concentration analyzer has achieved high-precision and stable ammonia gas analysis(NH3)Measurement. The analyzer has high sensitivity and a detection limit ofpptLevel, high stability, and negligible drift from one month of continuous work. Coating on important gas channels(SilcoNert ®)Can be reducedNH3The adsorption tendency of molecules on the channel surface improves the response time of measurements and eliminates measurement bias. Extra carbon dioxide(CO2)Measurement values are used forproxy authenticationSimplified and replaced the complex calibration process of standard gases (refer to page 2).
For ammonia gas that requires real-time response measurement (Figure)1)And/Or applications with high sensitivity and stability,PicarroAn analyzer is an ideal solution. These applications include but are not limited to urban and atmospheric air quality monitoring, particulate matter formation research, quantification of livestock emissions, quantification of vehicle emissions, indoor air quality, etc. The analyzer has a small size and low power consumption requirements, and can be disassembled and assembled within a few minutes (whether in the laboratory or in the field).
SI2103The analyzer can run for several months without user interaction, and the concentration trend data is continuously archived on the internal hard drive of the analyzer. And it can be achieved through EthernetRS-232Interface, analog signal4~20 mAperhapsModbusAutomatically export measurement data output.
PicarroCavity decay spectroscopy technology(CRDS)It combines many advantages, such as high precision, accuracy, low drift, and ease of use.
Ammonia response time

picture1 20 ppbUnder the change of ammonia concentration,SI2103 10~90%and90~10%Typical response time.
SI2103 Performance Specifications |
Ensure performance** |
Detection lower limit(3σ,100 Seconds) |
300 ppt |
zero drift*(50The difference between the maximum values of the minute averages |
± 100 ppt |
|
Accuracy(1σ,10 s) accuracy(1σ,100 s) |
≤300 ppt ≤100 ppt |
Measurement Interval*** |
< 4=''>s |
Response time(020 ppb)(Rising)/fall time10-90%/90-10%) |
< 1=''> |
measuring range |
0~10 ppm |
*The Picarro analyzer does not require zero point standard gas or zero point reagent kit to operate or meet specifications.
**When conducting span measurements, the measurement interval will be increased to twice the listed parameters.
***Each purchase of an analyzer comes with specifications and a test report (certificate of conformity) for the instrument.
SI2103 Proxy Gas Verification | |
For any gas phase analyzer, calibrating with ammonia standard gas is extremely challenging. Carbon dioxide(CO2)It is a commercially available gas with an absorption spectrum similar to that of ammonia, thereforeCan be used asverifymeasureExcellent proxy gas for accuracy and linearity. useCO2Okay, yesPicarro SI2103Verify without the need for useNH3Calibrate with standard gas.if requiredunderstandPicarroFor more information on proxy gas verification, please contactPicarroSales representative or application scientist:sales@picarro.com | |
SI2103 system specifications | |
measurement |
Cavity ring-down spectroscopy (CRDS) |
Measure chamber temperature and pressurestrongcontrol |
±0.005℃;±0.0002 atm |
sample temperature |
−10~45℃ |
Sample flow rate and pressure |
make an appointment2 slm(760support from underUnder atmospheric pressure);600~950support from underair pressure(80~127 kPa) |
Sample humidity |
<99%Relative humidity(40℃Under non condensing conditions,No need to dry) |
environmenttemperature |
10~35℃(When the instrument is working);−10~50℃(instrumentstoragetime) |
environmenthumidity |
< 85%relative humidity,Under non condensing conditions |
measureOther gases(Expected accuracy) |
H2O(< 200=''>,1σ,10 sec),CO2(< 10=''>,1σ,10 sec) |
join in marriagepiece |
wrapcontain:vacuumPump (external)place)、keyboard、mouse Optional: LCD display、16roadSilcoNert ®Coating sampler |
data output |
RS-232EthernetUSBAnalog signal0~10 V、Modbus、4~20 mA(Optional) |
Air inlet connector |
¼ inchSwagelok ®SSConnector (recommended ¼ inch)OD PFAPipe) |
shapesize |
Analyzer:43.2(Width) × 21.3(High) × 62(Long)cm, including foot pads External pump:19(Width) × 10.2(High) × 28(Long)cm |
installform |
Desktop or19Inch rack mounted chassis |
weight |
Analyzer < 25 kgThe external pump is6.5 kg |
Power requirements |
100~240 VAC power;47~63 Hz(Automatic detection); At startup (total)<375 W;Stable operationgotime:120 W(analyzer),150 W(Pump) |