SI2108 Hydrogen Chloride (HCl) Gas Concentration Analyzer

NegotiableUpdate on 02/24
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Measure HCl
Product Details

Hydrogen chlorideHCl)Gas concentration analyzer

Fast, continuous, and real-time measurement, without interference

high sensitivityAccuracy and precisionDegree, almost no drift

High dynamic range, high linearity

not haveNeed to useConsumables

It can be installed and put into use within a few minutes

● Sturdy and durable structure, insensitive to changes in environmental temperature

PicarroSI2108Hydrogen chloride(HCl)The analyzer canpptSensitivity for real-time measurement of trace amountsHClThe concentration. This analyzer is based onPicarroUnique cavity ring down spectroscopy technology(CRDS)The development is a system based on time measurement, which can use near-infrared laser to measure the spectral characteristics of molecules. Gas circulates in the optical cavity, with an effective optical path length of up to20 kmThe patented high-precision wavelength monitor ensures that the analyzer only monitors the spectral features of interest, greatly reducing the sensitivity of the analyzer to interfering gas components, and can measure ultra trace gas concentrations even in the presence of other gases. Therefore, the analyzer only requires minimal calibration work to maintain high linearity, accuracy, and precision under constantly changing environmental conditions.Picarro SI2108There is a precise chamber temperature and gas pressure control system inside, which can ensure accurate measurement for a long time while minimizing the use of calibration gases. This analyzer is easy to use, sturdy, virtually drift free, maintenance free, and requires almost no consumables.

This analyzer can be easily transported between stations, and can be set up and put into operation within minutes without the need to prepare any samples.

The gas concentration can be displayed in real-time without post-processing, and the data can be automatically and continuously archived to the internal hard drive of the analyzer.

This analyzer is designed to operate in laboratories and harsh environments, and can run for months without user interaction. Its built-in software includes a valve sequencer that can control up to6An external solenoid valve and1A rotary valve. This analyzer can be configured through EthernetRS-232Interface, analog signal4-20mAperhapsModbusAutomatically export measurement data using other methods.

Response time of hydrogen chloride analyzer

SI2108  氯化氢(HCl)气体浓度分析仪

picture1 answerSI2108The analyzer is20 ppbMultiple cycle response time tests conducted under changes in hydrogen chloride concentration. The fast gas response rate ensures that even short-term plumes can be detected.

SI2108 Performance Specifications*

HCl

H2O

Low detection limit(100 second

≤ 45 ppt

≤ 30 ppm

drift50The difference between the maximum values of the minute averages

± 50 ppt

± 40 ppt

Accuracy(1σ10 second

accuracy1σ100 Seconds)

45 ppt

15 ppt

20 ppm + 8 × H2O [%]

10 ppm + 4 × H2O [%]

Measurement Interval*

3 s

3 s

measuring range

0~2 ppm

0~40000 ppm

response time(Rising)/fall time10-90%/90-10%

1 ppm< 1=''>

10000 ppm< 10=''>s

*During span measurement, the measurement interval will be increased to twice the listed parameters

SI2108 system specifications

measurement

Cavity ring-down spectroscopy CRDS

Measure chamber temperature and pressurestrongcontrol

± 0.005℃± 0.0002 atm

sample temperature

−10~80℃

Sample flow rate and pressure

make an appointment2 slm600~950support from underair pressure80~127 kPa

Sample humidity

99%Relative humidity(40 ℃Under non condensing conditions, there is no need for drying

environmenttemperature

15~35℃Instrument operation);−10~50℃instrumentstoragecondition

environmenthumidity

85%relative humidityUnder non condensing conditions

Measureditsitgas

H2OCH4

join in marriagepiece

wrapcontainvacuumPump (external)placekeyboardmouse

Optional: LCD displayA0901

Operating system anddata output

RS-232EthernetUSBAnalog signal0~10 VModbus4~20 mA(Optional)

Air inlet connector

¼ inchSwagelok ®SSConnector (recommended ¼ inch)OD PFAPipe)

shapesize

Analyzer:43.2(Width) × 21.3(High) ×62(Long)cm, including foot pads

External pump:15.8(Width) × 22.6(High) × 32.4(Long)cm

installform

Desktop or19Inch rack mounted chassis

weight

Analyzer < 25 kgThe external pump is6.5 kg

Power requirements

100~240 VAC power;47~63 Hz(Automatic detection); At startup (total)<375WStable operationgotime120W(analyzer),150W(Pump)

authentication

CEauthentication