SI2205 Hydrogen Fluoride (HF) Gas Concentration Analyzer

NegotiableUpdate on 02/24
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Measure HF and H2O

Product Details

hydrogen fluorideHF)Gas concentration analyzer


Fast, continuous, real-time measurement without interference

pptLevel sensitivityAccuracy and precision, with almost no drift

Good long-term stability, no need for frequent calibration

Small size, can be deployed in the field or laboratory, no consumables required

● Automatic water calibration report for dry gas mole fraction

● Sturdy and durable structure, insensitive to changes in environmental temperature

PicarroSI2205Gas concentration analyzer canpptLevel sensitivity to hydrogen fluorideHFReal time and drift free measurement of concentration. This analyzer is based onPicarroUnique cavity ring down spectroscopy technology(CRDS)Developed to use near-infrared lasers to measure the spectral characteristics of molecules for time-based measurements. Gas circulates in the optical cavity, with an effective optical path length of up to20 kmThe patented high-precision wavelength monitor ensures that the analyzer only monitors the spectral features of interest, greatly reducing the sensitivity of the analyzer to interfering gas components, and can measure ultra trace gas concentrations even in the presence of other gases. Therefore, the analyzer only requires minimal calibration work to maintain high linearity, accuracy, and precision under constantly changing environmental conditions.Picarro SI2205There is a precise chamber temperature and gas pressure control system inside, which can ensure measurement accuracy for a long time while minimizing the use of calibration gases. This analyzer is easy to use, sturdy, virtually drift free, maintenance free, and requires almost no consumables. Therefore, the device has excellent usability and effectively reduces the cost of subsequent long-term measurements.

This analyzer can be easily transported between stations, set up and put into operation within minutes, and does not require any sample preparation. Its built-in valve sequencer software can control up to6An external solenoid valve and1A rotary valve. The gas concentration can be displayed in real-time without post-processing, and the measurement data can be automatically and continuously archived to the internal hard drive of the analyzer. This analyzer is designed to operate in laboratories or harsh environments and can run for months without user interaction. This analyzer can be configured through EthernetRS-232InterfaceAnalog 4-20mAperhapsModbusAutomatically export measurement data using other methods. Users can remotely connect to the analyzer through standard remote desktop connections or by using similar remote login softwareLinuxSystem.

Hydrogen fluoride response time

SI2205 氟化氢(HF)气体浓度分析仪

picture1 PicarroSI2205HFQuick response of analyzer to changes in measured concentration

SI2205 sexualCapable of specifications*

HF

accuracy100 s

10 ppt

limit of detection100 s

30 ppt

drift50The difference between the maximum values of the minute averages

± 25 ppt

measuring range

0~1 ppm

Measurement Interval

4 s

Response time(0 – 20ppb(Rising)/fall time10-90%/90-10%

< 1=''>

*After 1 hour of stable equilibrium, ensure the following operating conditions are met.

SI2205 system specifications

measurement

Cavity ring-down spectroscopy CRDS

Measure chamber temperature and pressurestrongcontrol

± 0.005 ℃± 0.0002 atm

sample temperature

−10~45 ℃

Sample flow rate and pressure

make an appointment2 slm760support from underUnder atmospheric pressureNo need for filtering;600~950support from underair pressure80~127 kPa

Maximum rate of change in ambient temperature

5 ℃/hr

Sample humidity

relative humidity99%40℃Under non condensing conditions.H2OIt will affectHFMeasurement. To achieve optimal performance, you can choose toHFDry the gas or maintain consistency during measurementH2Oconcentration

environmenttemperature

10~35 ℃Instrument operation);−10~50 ℃instrumentstoragecondition

environmenthumidity

85%relative humidityUnder non condensing conditions

Measureditsitgas

H2OO2

join in marriagepiece

wrapcontainvacuumPump (external)placekeyboardmouse

Optional: LCD displayA0901)Rack mounted(A0954

data output

RS-232EthernetUSBAnalog signal0~10 VModbus4~20 mA(Optional)

Air inlet connector

¼ inchSwagelok ®SSConnector (recommended ¼ inch)OD PFAPipe)

shapesize

Analyzer: 43.2(Width) × 21.3(High) × 62(Long)cm, including foot pads

External pump:15.5(Width) × 22(High) × 34.5(Long)cm

installform

Desktop or19Inch rack mounted chassis

weight

Analyzer < 25 kgThe external pump is6.5 kg

Power requirements

100~240 VAC power;47~63 Hz(Automatic detection); At startup (total)<375W; StableFixed runtime120W(analyzer),150W(Pump)

authentication

CEauthentication

Application precautions

The calibration of this instrument is limited by the preparation of constant concentration samples. We should investigate whether the instrument is suitable for commercial standards based on your application. For more details, please contactPicarro.