Welcome Customer !

Membership

Help

Chongqing Sainz Instrument Co., Ltd
Custom manufacturer

Main Products:

pharmamach>Products

SS705 online multi parameter water quality monitoring instrument

NegotiableUpdate on 04/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

SS705 online multi parameter water quality monitoring instrument is suitable for environmental protection, agriculture, manufacturing, education, scientific research, and comprehensive service industries to conduct multi parameter analysis of water quality

Product Details

Online multi parameter water quality monitoring instrumentTechnical parameters

1. temperatureMeasurement range: (-10.0~120.0) ℃

Basic error of electronic unit: ± 0.3 ℃ ± 1 word

Basic instrument error: ± 0.5 ℃± 1 word (0.0 ℃≤ T ≤ 60.0 ℃);

Other range ± 1.0 ℃ ± 1 word

2. PHMeasurement range: (0.00~14.00)pH

Basic error of electronic unit: ± 0.03pH ± 1 word

Basic error of the instrument: ± 0.05pH ± 1 word

3. ORPMeasurement range: (-1999~1999)mV

Basic error of electronic unit: ± 1mV ± 1 word

Basic error of the instrument: ± 10mV ± 1 word

4. Conductivity/TDS

Conductivity measurement range: 0.000 μ S/cm to 0.199 μ S/cm

0.20 μS/cm~1.99 μS/cm

2.0 μS/cm~19.9 μS/cm

20 μS/cm~199 μS/cm

0.2 mS/cm~1.99 mS/cm

2.0 mS/cm~19.9 mS/cm

TDS measurement range: 0.000 mg/L to 0.099 mg/L

0.10 mg/L~0.99 mg/L

1.0 mg/L~9.9 mg/L

10 mg/L~99 mg/L

0.10 g/L~0.99 g/L

1.0 g/L~9.9 g/L

Basic error of electronic unit: ± 1.0% (FS)

Basic instrument error: ± 2.0% (FS)

5. Dissolved oxygen concentration/saturation

Measurement range: Dissolved oxygen concentration: (0.00~20.00)mg/L;

Dissolved oxygen saturation: (0.0~200.0)%

Basic error of electronic unit: dissolved oxygen concentration: ± 0.20mg/L;

Dissolved oxygen saturation: ± 5% (FS).

Basic instrument error: Dissolved oxygen concentration: ± 0.50mg/L when the temperature difference between the calibration water sample and the measured water should not exceed ± 15.0 ℃;

Dissolved oxygen saturation: ± 10.0% when the temperature difference between the calibration water sample and the measured water does not exceed ± 15.0 ℃.

6. Normal working conditions of the instrument:

Environmental temperature: (5.0-35.0) ℃;

Relative humidity: not exceeding 85%;

Power supply: AC voltage (220 ± 22) V;

Frequency (50 ± 1) Hz;

There are vibrations around that do not affect performance;

There are no corrosive gases present in the surrounding air;

Electromagnetic field interference that does not affect performance other than the geomagnetic field in the surrounding area

7. Dimensions: 600mm * 455mm * 720mm

Online multi parameter water quality monitoring instrumentProduct Features