Overview
The SMPSTM scanning electron mobility particle size spectrometer launched by TSI is widely used as a measurement standard for measuring the particle size distribution of aerosols below 1 micron. This system is also applied to accurately measure the nanoparticle size of liquid suspended particles. The National Institute of Standards and Technology (NIST) in the United States measured the standard reference particle sizes of 60nm and 100nm using a TSI differential electrostatic mobility analyzer (DMA). SMPSTM scanning electron mobility particle size spectrometer particle size measurement is a cautious technique, in which the counting concentration is directly measured without assuming the shape of the particle size distribution.
Product Details

3757 type nanoenhancer

particle size range

Minimum detectable

Particle (D50): 1.4 nm (electric mobility diameter, geometric diameter 1.1 nm), verified with sodium chloride particles


traffic

Aerosol flow rate: 2.5 L/min

Aerosol outlet flow rate: 1.0 L/min

Transport flow rate: 1.5 L/min

Flow source: external vacuum

Flow control: The volumetric flow is controlled by the internal orifice plate of the conveying flow. The aerosol flow rate is controlled by the 3750 CPC.


Aerosol medium

Suggest using it together with air; Can be safely used with inert gases such as nitrogen, argon, and helium (technical parameters applicable to air)


Liquid system

Diethylene glycol (DEG, ≥ 99%, exclusive) is used as the working fluid. Use a water separator and a refillable dryer to dry the sheath gas. Suggest using an internal drainage pump to remove condensed water.


Communication

Embedded touch display screen, USB-C type, can directly connect the nanoenhancer to the 3750 CPC.


Required attachments

Electrical: 100 to 240 VAC, 50/60 Hz, minimum power 240W. Automatic recovery of built-in power failure

Vacuum: Minimum pressure of 60 kPa (18 in Hg) (below atmospheric pressure)

Includes: water injection and drainage bottles, dryer


Physical characteristics

Size: (H x W x D) 30x28.2x32.5cm (11.8x11x12.6in.), excluding water bottle and holder

Weight: 9.1 kg (20 lbs)

1nm CPC system (3757 nanoenhancer and 3750 CPC)

particle size range

Minimum detectable

Particle (D50): 1.4 nm (electric mobility diameter, geometric diameter 1.1 nm), verified with sodium chloride particles

Particle concentration range

0-3x105 particles/cm3, single particle continuous counting, real-time consistency correction.

Particle concentration accuracy

When<1.65 x 105 particles/cm3, it is ± 10%, and when 3 x 105 particles/cm3, it is ± 15%

Error counting background

<0.01 x 105 particles/cm3, based on a 12 hour average

response time

<4s 95% - response to significant changes in concentration

Catalog Items

Model Description

3757-50 1 nm condensed nucleus particle counter

3032 vacuum pump 110V

3032-1 Vacuum Pump 230 V/50 Hz

3750-MKIT CPC Repair Tool

3750-WKIT CPC Suction Core Replacement Tool

3772200 Environmental Sampling System


Scanning electric mobility particle size spectrometer

Model: 3938


TSI3938, the SMPS TM scanning electron mobility particle size spectrometer launched by TSI, is widely used as a measurement standard for measuring the particle size distribution of aerosols below 1 micron. This system is also applied to accurately measure the nanoparticle size of liquid suspended particles. The National Institute of Standards and Technology (NIST) in the United States measured the standard reference particle sizes of 60nm and 100nm using a TSI differential electrostatic mobility analyzer (DMA). The SMPS TM scanning electron mobility particle size spectrometer is a cautious technique for particle size measurement, in which the counting concentration is directly measured without assuming the shape of the particle size distribution. This method does not rely on the refractive index of particles or liquids and has high absolute particle size measurement accuracy and measurement repeatability. The third generation 3938 SMPS adopts a modular, component-based design, which not only has a very high particle size resolution, but also has fast scanning (less than 15 seconds per scan) and automatic component recognition function. TSI SMPS is easy to use and can provide researchers with the highest quality data, making it the best choice for measuring nanoparticle size distribution. TSI's 3938 instrument is the third generation product of SMPS and has been trusted by researchers for 30 years.


application

·Nanotechnology research and material synthesis

·Atmospheric research and environmental monitoring

·Research on Engineering Nanomaterials Science

·Filter and air purifier testing

·Research on Combustion and Engine Emissions

·Indoor air quality measurement

·Research on nucleation/condensation

·Inhalation toxicology research

including components

·Static electricity classifier compatible with your self selected DMA differential electrostatic mobility analyzer

·One of the seven condensed particle counters

· Aerosol Instrument Manager ® Aerosol instrument management software

The computer used for data collection must be purchased separately.


Function:

The SMPS particle size spectrometer consists of the following components: 3082 electrostatic classifier, 3086 1nm DMA, 3757 nanoenhancer, and 3750 condensed core particle counter

+High resolution particle size distribution;

+64 channels per 10 times particle size

+The number of channels between 1-50nm is greater than 109

+Component design with maximum flexibility

+Wide particle size range of 1nm-50nm

+By being used in conjunction with the 3081a long DMA, it is possible to measure the particle size distribution of three 10 fold sizes ranging from 1nm to 1 µ m

+Optimize for minimum diffusion loss and system integration

+Use the Aerosol Instrument Manager (AIM) software for unified system operation

+Precise particle measurement: suitable for multimodal samples

Remarks: TSI Particle Size Spectrometer 3938, TSI Scanning Electron Mobility Particle Size Spectrometer 3938, TSI3938, acting as the agent for TSI Scanning Electron Mobility Particle Size Spectrometer 3938, TSI3938, TSI Particle Size Spectrometer, TSI Particle Imaging Instrument.