Shanghai gas chromatograph manufacturer provides industrial propylene detection

NegotiableUpdate on 04/02
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Design a timed self starting program that can easily complete online analysis of gas and liquid samples (with injection components required);
Product Details

Shanghai gas chromatograph manufacturer provides industrial propylene detection

Instrument features:

★ Adopting a 5.7-inch industrial color LCD screen design, the display content is comprehensive, and the interface operation is more convenient;

★ The display interface is interchangeable between Chinese and English to meet the needs of different users;

★ Adopting mold injection molded buttons, beautiful, atmospheric, and durable;

Multiple instruments can be combined to achieve multi-dimensional chromatographic analysis of multiple data simultaneously through a single injection.

★ Advanced 10/100M adaptive Ethernet communication interface and built-in IP protocol stack are adopted, which is convenient for enterprises to realize long-distance data transmission through internal LAN and Internet; Facilitating laboratory setup, simplifying laboratory configuration and data management;

★ Equipped with NETChrom ® The workstation can support N chromatographs to work simultaneously, achieving data processing and reverse control, reaching the level of the industry;

★ Internally designed with 3 independent connection threads, which can connect to local processing, unit supervisors (such as chief engineer, technical factory director, etc.), and higher-level supervisory departments (such as environmental protection bureau, technical supervision bureau, etc.), facilitating real-time monitoring of instrument operation and analysis of data results by unit supervisors and higher-level supervisory units;

★ Adopting a modular structural design, the design is clear and easy to replace and upgrade, protecting the effectiveness of investment and meeting the requirements of complex sample analysis. It can be equipped with 7 high-performance detectors, such as FID, TCD, ECD, FPD, NPD, PID, and PDDID, etc;

★ NETChrom ® The Modbus/TCP server built into the workstation allows for easy integration of analysis results into DCS (Distributed Control System);

★ * Abandoning traditional pointer pressure gauges and being able to load EPC technology for gas path control, the automation level and overall performance are close to *;

★ It has achieved self-protection against air circuit faults, automatic ignition and shutdown, and automatic opening of the air circuit, achieving one key start;

★ Design a timed self starting program that can easily complete online analysis of gas and liquid samples (with injection components required);

The machine has an automatic sampler interface and can be equipped with multiple automatic samplers; Can interface with AOC-20i *;


Chromatographic conditions:

Chromatographic column: 30 * 0.32 * 25

Column temperature: 120 ℃

Carrier gas: N2 30min/sec

Injector: 200 ℃ split injection

Detector: FID 200 ℃

Sample analysis: Industrial propylene

Shanghai gas chromatograph manufacturer provides industrial propylene testingmeasureMain technical indicators:

● Interface display: 5.7-inch industrial color LCD screen

Temperature control area: 8 channels, can control dual column boxes, each column box has an independent rear door design.

Temperature control range: 4 ℃ to 450 ℃ above room temperature, increment: 1 ℃, accuracy: ± 0.01 ℃

● Program heating order: 16 orders, 17 platforms

● Lift rate: 0.1-39 ℃/min (ordinary type); 0.1-80 ℃/min (high-speed type)

External events: 8 channels; The auxiliary control output consists of 2 channels

● Sample injector types: packed column injection, capillary injection, gas valve injection, headspace injection, thermal analysis injection, blowdown collector injection, flash evaporation injection, etc

● Number of detectors: up to 9 high-performance detector options are available, including FID, TCD, ECD, FPD, NPD, PID, PDI ID, ZD, DID, etc., and 3 types of detectors can be installed simultaneously

● Gas path control: Mechanical valve control mode and EPC mode can be selected, or the two modes can be mixed and used.

EPC and EFC working modes: 2 types; Constant current mode, constant voltage mode

EPC and EFC working gases: 5 types; Nitrogen, hydrogen, air, helium, argon

EPC and EFC lift: 4th order

EPC and EFC control range: pressure: 0-0.6MPa; Flow rate 0-100mL/min or 0-500mL/min (air)

● Pressure sensor:

Accuracy:<± 2% of full scale

Reproducibility:<± 0.05 KPa

Temperature coefficient:<± 0.01 KPa/° C

● Flow sensor:

Accuracy:<± 5% of full scale

Reproducibility:<± 0.5% (full scale)

Range: 0-500mL/min

● Start injection: manual or automatic selection

● Communication interface: Ethernet: IEEE802.3

Power supply: 220V ± 10%, 50Hz; 2500W (maximum)

● Volume: 572 × 552 × 465 (height) mm

Weight: 50kg (approximately)

Technical specifications of detector

Hydrogen flame ionization detector (FID)

★ Detection limit: Mt ≤ 3 × 10-12g/s (n-hexadecane-isooctane solution);

★ Baseline noise: ≤ 5 × 10-14A

★ Baseline drift: ≤ 1 × 10-13A/30min

★ Linear range: ≥ 106

Thermal conductivity detector (TCD):

★ Sensitivity: S ≥ 3500mV • ml/mg (benzene solution) (standard), maximum sensitivity 10000mV • ml/mg (equipped with electronic amplifier board, 2, 4, 8 times optional)

★ Baseline noise: ≤ 10 μ V

★ Baseline drift: ≤ 30 μ V/30min

★ Linear range: ≥ 104

Electron Capture Detector (ECD):

★ Detection limit: ≤ 1 × 10-14g/s

★ Baseline noise: ≤ 3 × 10-13A

★ Baseline drift: ≤ 0.2mV/30min

★ Linear range: ≥ 103

★ Radioactive source: 63Ni

Flame photometric detector (FPD):

★ Detection limit: (S) ≤ 5 × 10-11g/s, (P) ≤ 1 × 10-12g/s;

★ Baseline noise: ≤ 3 × 10-13A

★ Baseline drift: ≤ 0.2mV/30min

★ Linear range: ≥ 103 (S), 102 (P)

Helium ionization detector (PDIHID):

★ Discharge mode: Pulse discharge

★ Baseline noise: ≤ 0.1mv

★ Baseline drift: ≤ 0.5mv/30min

★ Detection limit: ≤ 1.0 × 10-10g/ml

Photoionization detector (PID)

★ Photon energy 10.6eV

★ Light source lifespan ≥ 3000h

★ Detection limit: ≤ 1ug/m3 (benzene)

★ Baseline noise: ≤ 8 × 10-14A

★ Linear range: ≥ 105

Determination of impurities in propylene

Chromatographic column: 30 * 0.32 * 25

Column temperature: 120 ℃

Carrier gas: N2 30min/sec

Injector: 200 ℃ split injection

Detector: FID 200 ℃

Sample analysis: Industrial propylene