| Main features: | |
| -Fully intelligent design, fully automated system truly achieves off machine analysis. | |
| -The design of the touch screen can display real-time instrument pipeline diagrams, valve actions, sensor readings, isotherms, and ongoing status, as well as be used to set specific surface area and aperture analysis methods and diagnostic instruments. | |
| -Compatible with multiple gas analyses without the need for attachments, nitrogen, argon, and krypton can be used for adsorption analysis as needed. | |
| -The unique "concurrent gas injection" technology fully improves analysis efficiency and saves analysis time. This technology utilizes the equilibrium time during the sample adsorption process, while balancing the air pressure in the manifold, so that the next gas injection and adsorption process can proceed sequentially. Because the amount of adsorbed gas each time needs to be calculated using the volume of the manifold and the equilibrium pressure. | |
| -The isotherm memory function makes the analysis of similar samples extremely fast. Being able to remember the analyzed isotherms allows the instrument to refer to the most suitable gas injection program when reanalyzing similar samples, achieving the goal of saving time. And improve the efficiency of quality control analysis work. | |
| -The static gas injection method is used to provide sufficient equilibrium time for the adsorption process of the sample, ensuring the accuracy of the specific surface area and pore size analysis results. | |
| -The rapid helium dead space calibration technology makes the analysis results more reliable and accurate (especially for samples with small specific surface areas). | |
| -In response to the instability of the reference tube, the instrument is specifically designed to measure the saturation pressure at each time the sample pressure data is obtained, in order to ensure the accuracy of the relative pressure (sample pressure/saturation pressure) data for each group. | |
| -Powerful and flexible software controls all functions, and can display pressure readings and isotherms in real-time during the analysis process. | |
| -SA View software and Sorption Analysis software provide you with various functions such as data storage, comparison, and printing. | |
| -The Sorption Analysis software has multiple user identity selection and a four level data security system to ensure the authenticity of data. The software complies with the electronic record and electronic signature guidelines (21 CFR Part 11) formulated by the FDA, and the recorded data can be converted into files in different formats for easy communication of analysis results. | |
SA3100 Technical Parameters: | |
| -Analysis scope: | |
| Specific surface area: 0.01-2000m2/g or more. If Kr is used as the adsorbed gas, the lower limit can reach 0.001m2/g | |
| Pore size distribution: 20 Å -2000 Å | |
| -Analysis functions: single point and multi-point BET, Langmuir specific surface area, BJH adsorption and desorption, pore volume distribution, pore area distribution, total pore volume, T-graph, etc | |
| -Repetitive error:< 1% | |
| -Analysis time: 15-30 minutes (typical materials) | |
| -Pipeline temperature: 45 ℃~+/-0.1 ℃ | |
| -Heating furnace: | |
| Temperature range: 30 ℃ -350 ℃; Adjustable: 1 ℃; Stability: 5 ℃; Accuracy:+/- 5℃ | |
| -Pressure control: | |
| Vacuum degree: 10-3-10-4Torr | |
| Zui small relative pressure: 6 × 10-5 | |
| Resolution: 0.046mmHg | |
| Sensor linearity:< 0.1% BFSL | |
| Sensor repeatability: 0.0125% BFSL | |
| -Vacuum pump: | |
| Extraction volume velocity: 1.5m3/hr | |
| Power consumption: 130W | |
| Zui High Vacuum: 10-5Torr | |
| -Operating conditions: | |
| Power: 500W | |
| Operating temperature: 10℃-35℃ | |
| Relative humidity: 10% -85% | |
| -Gas requirements: | |
| He: 99.995%; N2: 99.9%; N2 (liquid): 99.9%; Kr: 99.99%; Ar: 99.99% | |