- Phone
-
Address
2nd Floor East, Building A2, Shunheda Industrial Park, Liuxiandong Industrial Zone, Xili Street, Nanshan District, Shenzhen
Shenzhen Changhong Technology Co., Ltd
2nd Floor East, Building A2, Shunheda Industrial Park, Liuxiandong Industrial Zone, Xili Street, Nanshan District, Shenzhen


|
【 Host 】 | |||
| Measurement range: | 0.10~150mg/L | Indication error: | ≤±5% |
| Repeatability: | ≤3% | Display: | Color touch screen, displaying in both Chinese and English |
| Temperature control system: | The room temperature can be set between 200 ℃, and the digestion temperature is 125 ℃. | Optical stability: | The absorbance value of the instrument drifts less than 0.002A within 20 minutes |
| Temperature control accuracy: | ±0.5℃ | Temperature control time: | 1~9999min adjustable |
| Dissolving time: | 30 min | Batch processing capacity: | 16 water samples |
| External dimensions: | Measuring instrument 400mm × 310 mm × 158mm; digestion instrument (DIS-16B type) 355mm × 260mm × 135mm | ||
| Weight: | Host 4kg Dissolver 5.5kg | power waste: | <500W |
| Normal usage conditions: | |||
| Environmental temperature: | 5~40℃ | Relative humidity: | ≤85% |
| Power supply: | AC(220±22)V;(50±0.5)Hz | No significant vibration or electromagnetic interference, avoid direct sunlight. | |
|
The measuring instrument is a cold light source and narrowband interferometric optical system with good optical stability. Both the measuring instrument and the digestion instrument are color touch screens with a user-friendly interface and display in both Chinese and English. Simple and time-saving operation. Dissolving colorimetric does not require changing the tube. The digestion instrument has a timed reminder and can be set to automatically shut down after digestion. The instrument has a secondary protection function. When the temperature of the digestion instrument exceeds 200 degrees, the instrument will cut off the power by itself to ensure the safety of the digestion process. The digestion instrument has multi-point error free calibration, achieving zero error measurement of the set point, completely eliminating the errors caused by the nonlinearity of temperature sensors in common digestion instruments on the market. |
![]() It can store 300 standard curves and 1000 measured values without losing them when powered off. (Type D consists of 3500 curves and 100000 data points) It is possible to query the curve, in addition to querying the curve equation, it is also possible to query the standard value and corresponding absorbance of each standard substance during calibration, as well as the calibration time and calibration personnel number, fully reproducing the state during calibration. Optical system optimization upgrade (D-type), the system is more stable during measurement, with better accuracy and repeatability. |
![]() Printing function: It can print single items or pages of test records. USB interface, can be connected to a computer. |
![]() The instrument has a one click recovery function to prevent accidental loss of curves and data. The instrument comes with a calibration function that allows for self calibration during operation, effectively eliminating drift errors caused by long-term use. |
| name | quantity | name | quantity |
| host | 1 unit | Dissolving colorimetric tube | 10 pieces |
| Digester | 1 unit | Dissolving tube | 10 pieces |
| test tube rack | 1 piece | Total nitrogen reagent | 1 set |
| Dissolve the protective cover | 1 piece | UBS data | 1 piece |
| Data collection CD | 1 sheet | operation instructions | 1 copy |
| Product Qualification Certificate | 1 copy | warranty card | 1 copy |