E-mail
671113565@qq.com
-
Phone
13761235453
-
Address
No. 265 East Street, Zhuangxing Town, Fengxian District, Shanghai
Shanghai Chengsi Intelligent Technology Co., Ltd
671113565@qq.com
13761235453
No. 265 East Street, Zhuangxing Town, Fengxian District, Shanghai
Thermal conductivity tester - transient plane heat source method
CSI-L007 Transient Plane Heat Source Testing Machine、Multi functional rapid thermal conductivity tester、Thermal conductivity tester-Transient plane source method
1、 Description:
This instrument adopts the transient plane heat source method and tests materials based on the transient plane heat source technology: thermal conductivity coefficient, thermal conductivity coefficient (thermal diffusion coefficient), specific heat capacity, thermal storage coefficient and thermal resistance of solid, powder, liquid, paste, coating, film, thermal insulation material, thermal insulation material, anisotropic material, such as metal, alloy, graphite, thermal conductive silicone grease, silica gel, silicone rubber, ceramics, rock, polymer, paper, fabric, foamed plastic, concrete, composite plate, paper honeycomb plate, etc.
2、 Execution standards
Standard:ISO 22007-2:2008, GB/T32064-2015
IIITechnical Specifications:
Control system:PLC+Windows system;
Operating interface: Color12 inch embedded industrial computer, WiFi connection;
Wireless printing function,Using fully automated testing software to quickly and accurately analyze the parameters of the sample during the testing process and output reports;
Measurement range of thermal conductivity:0.001~500 w/(m.k), The resolution is 0.0001w/(m.k);
Range of thermal diffusivity (thermal conductivity):0.1~100 m2/s;
Range of heat storage coefficient:0.1~30 w/(m2.k);
Specific heat capacity range:0.1~5 kJ/(kg.℃);
Thermal resistance range:0.5~0.000005 mm2.k/w;
Temperature measurement accuracy:≤0.001℃;

Measurement relative error:≤3%;
Repetitive error:≤ 3%;
Test time:1-120 seconds.
Temperature range of sample testing environment:optional-40~150℃; Standard configuration: Room temperature.
Probe configuration: Standard configuration: DiameterΦ15mm。
Optional probe diameter:Φ30mm, Φ15mm,Φ7.5mm;
power supplyAC 220V±10%,50/60 Hz;
Whole machine power: <500w;
Standard external dimensions:L761* W500* H786mm;
Optional external dimensions: Long921 wide, 600 high, 856.5mm
4Probe technical specifications:
1Heating material: nickel metal;
2Operating temperature:-50~150℃;
3Zui High Power:20w;
4、zuiHigh voltage:20V;
5Zui high current:1A;
6Resistance:About 10 Ω, according to the resistance value marked on the probe;
7Nickel wire specification: thickness0.01mm, Hot wire width: 0.35 ± 0.03mm;
8Hot wire spacing:0.35±0.03mm ;
9Protective layer material: polyimide film; Single layer thickness:0.06mm;
10Probe diameter:Φ30mm,Φ15mm,Φ7.5mm。
11Total thickness of probe (including adhesive layer thickness)0.07±0.02mm。