Welcome Customer !

Membership

Help

Guangdong Yuelian Instrument Co., Ltd
Custom manufacturer

Main Products:

pharmamach>Products

YLP-2000-60M High Low Temperature and Low Pressure Test Chamber

NegotiableUpdate on 10/31
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

YLP-2000-60M High Low Temperature and Low Pressure Test Chamber

Product Details

  • Product Application

    This product is manufactured according to user requirements, referring to the technical conditions of GB10590 low temperature/low pressure test chamber, GB 10589 low temperature test chamber, and GB 11158 high temperature test chamber. It mainly provides temperature and humidity change environments for scientific research and production units such as aerospace, aviation, petroleum, chemical, military, automotive, marine, electronics, and communication, simulates the adaptability test of test samples under temperature and humidity change environmental conditions, and provides reliability testing, product screening, and safety testing for electronic components; Used for high temperature testing, low temperature testing, and temperature low pressure testing functions for various connectors, cables, micro motors, etc.

    Product Features

    The experimental system has an advanced and reasonable structural design, standardized manufacturing processes, and a beautiful and elegant appearance.

    The main functional components of this test chamber are equipped with world-renowned brand configurations (high gold content), advanced and reliable technical principles, and optimal control of noise and energy saving - its performance can be comparable to similar foreign products.

    The matching and assembly of components are good, and the main functional components are all imported original parts with international advanced level, which improves the safety and reliability of the product and can ensure the long-term and high-frequency use requirements of users.

    The equipment has good operability, maintainability, good temperature stability and durability, good safety performance, does not pollute the environment and harm personal health;

    Communication protocol: Provides programming interface functions and requires monitoring software; Curve recording function: Equipped with RAM with battery protection, it can save the device's set values, sampling values, and sampling time. Display: TFT color LCD display; Operation mode: program mode, fixed value mode; Setting method: Chinese English menu

    Technical Specifications

    temperature range

    -60~+100℃

    temperature fluctuation

    temperature deviation

    Temperature Uniformity

    ≤± 0.5 ℃ (normal pressure, no load)

    ≤± 2.0 ℃ (normal pressure, no load)

    ≤ 2.0 ℃ (normal pressure, no load)When the device reaches the set temperature 1 Measure after hours

    heating rate

    cooling rate

    -70 ℃ →+150 ℃ ≤ 60min (no load, normal pressure, average throughout the process)(Mechanical refrigeration, under standard load)

    +25 ℃ → -20 ℃ ≤ 28min (no load, normal pressure, average throughout the entire process)

    +30 ℃ → -40 ℃ ≤ 45 minutes (no load, normal pressure, average throughout the process)

    +30 ℃ → -60 ℃ ≤ 80min (no load, normal pressure, average throughout the entire process)


    pressure range

    Atmospheric pressure ranging from 100kPa to 1kPa


    Voltage reduction rate

    Normal pressure~1kPa ≤ 30min (dry inside the box)

    At ambient temperature and in the working space of the drying box, within less than 5 minutes, it can be compressed from the environment

    The force drops to a height of 10000 meters (approximately 260 mBar); Within less than 25 minutes, it is possible to

    Reduced from environmental pressure to a height of 30500 meters (approximately 11 mBar)The pressure change rate is adjustable and can meet the requirement of not exceeding 100mBar/min (10kPa/min)速率要求


    pressure deviation

    ±2kPa (≥40kPa)

    ≤±5% (4~40kPa)

    ≤±0.1kPa(≤4kPa)


    Pressure holding time

    Not less than 10 hours

    Corresponding range of pressure

    Set pressure to correspond to a height of 0.5 kPa43000m1.09kPa30500m2.75kPa24400m4.43kPa20350m11.68kPa15250m19.16kPa12200m30.06kPa9150m46.54kPa6100m57.3kPa4550m69.66kPa3050m

    riseVoltage reduction rate

    Heating and cooling rate

    Pressure rise and fall rate control (average)

    Program control: altitude change rate 25m/min~500m/min

    Temperature rise and fall rate control

    Program control: ≥ 2 ℃/min

    Temperature uniformity: ≤± 2 ℃

    Temperature fluctuation: ≤± 1 ℃

    Setting and Adjusting Pressure

    section

    Pressure can be preset, pressure resolution: 0.01kPa

    The adjustable range for the rate of increase and decrease is 0.1kPa/min to 3.4kPa/min;

    Adjust the resolution to 0.1kPa

    Setting and Adjusting Pressure

    section

    Pressure can be preset, pressure resolution: 0.01kPa

    The adjustable range for the rate of increase and decrease is 0.1kPa/min to 3.4kPa/min;

    Adjust the resolution to 0.1kPa