High pressure vessel that meets the experimental requirements of the 2020 Chinese Pharmacopoeia for drug packaging material testing method
Date: 2022-11-25Read: 8

Product Name: Glass Water Resistance Test High Pressure Reactor

型 号: YXBN-53L,YXBN-70L,YXBN-100L,YXBN-120L。

Product Features:

1. Wet heat sterilization, conventional sterilization of culture media, agarose, glassware, clothing, etc.

2. Glass Particle Water Resistance Test (4001)

Place the conical flask into a pressure steam sterilizer, open the exhaust valve, and heat it uniformly. After 20-30 minutes, allow a large amount of steam to escape from the exhaust port and continue to escape for 10 minutes. Close the exhaust valve and continue heating, raising the temperature to 121 ℃ at an average rate of 1 ℃/min within 20-22 minutes Start timing when the temperature is reached, maintain at 121 ℃ for 30 minutes ± 1 minute, slowly cool and depressurize, and reduce the temperature to 100 ℃ at an average rate of 0.5 ℃/min within 40-44 minutes (to prevent vacuum formation). When the temperature is below 95 ℃, remove it from the pressure steam sterilizer and cool it to room temperature. (The product meets the experimental requirements of the 2020 edition of the Chinese Pharmacopoeia for the detection of pharmaceutical packaging materials)

3. Glass inner surface water resistance test (4006)

Place the test sample in a pressure steam sterilizer, open the exhaust valve, and heat it at a constant speed. After 20-30 minutes, allow a large amount of steam to escape from the exhaust port and continue to escape for 10 minutes. Close the exhaust valve and continue heating at an average rate of 1 ℃/min to raise the temperature to 121 ℃ within 20-22 minutes Start timing when the temperature is reached, maintain at 121 ℃ for 60 minutes ± 1 minute, slowly cool and depressurize, and reduce the temperature to 100 ℃ at an average rate of 0.5 ℃/min within 40-44 minutes (to prevent vacuum formation). When the temperature is below 95 ℃, remove it from the pressure steam sterilizer and cool it to room temperature. (The product meets the experimental requirements of the 2020 edition of the Chinese Pharmacopoeia for the detection of pharmaceutical packaging materials)

4. Aluminum Cover Coating Firmness Test Procedure (YBB00372003-2015)

Take an appropriate amount of this product and hold it at 121 ℃± 2 ℃ for 30 minutes, including steam sterilization at 130 ℃± 2 ℃ for 5 minutes. (The product meets the experimental requirements of the 2020 edition of the Chinese Pharmacopoeia for the detection of pharmaceutical packaging materials)

5. Puncture force of rubber plug gasket (4015)

Satisfy the first method, maintain 121 ℃ for 20 minutes,

The second method is to heat up to 121 ℃ within 30 minutes and cool down to room temperature within 30 minutes. Special customization is required for rapid heating and cooling. Note: Before customization, it is necessary to confirm the loading quantity and type of items.

6. Rubber plug gasket puncture debris (4016)

Satisfy the first method, maintain 121 ℃ for 20 minutes,

The second method is to heat up to 121 ℃ within 30 minutes and cool down to room temperature within 30 minutes. Special customization is required for rapid heating and cooling. Note: Before customization, it is necessary to confirm the loading quantity and type of items.