Precautions for Light Stability Test Chamber
Date: 2024-11-19Read: 8
The light stability test chamber is equipped with visible light and near ultraviolet lamps, which can independently control the type of light source and can print and record visible light intensity and near ultraviolet radiation in real time; Visible light and near ultraviolet can be directly set, automatically adjusted, and accurately controlled. The test chamber is used for research experiments in the pharmaceutical industry, medicine, biotechnology industry, and all related industries including life sciences.
Precautions for Light Stability Test Chamber:
1. Do not approach walls or other objects around the device, and leave a certain gap (greater than 0.9 meters in the front and greater than 0.3 meters in the left and right rear) to facilitate heat dissipation.
2. The operating environment temperature of the equipment is 5-35 ℃, and the humidity is not much than 85%.
3. When the equipment is not in use for a long time, it is necessary to remove the internal moisture, set the temperature to 40 ℃ and the humidity to 0%, and turn it on for 3-4 hours. After running, the box door should be opened to facilitate the removal of moisture.
4. About six months later, the air inlet of the compressor needs to be dedusted.
5. Before starting up, there must be water in the heating water tank to submerge the heating tube. If not, use a beaker to add water to the overflow port.
6. Remember to maintain the water level in the storage tank, which should not be lower than the water level at the interface of the submersible pump, to prevent dry burning and damage to the submersible pump and heating pipe.
7. Due to the long-term growth of microorganisms in the water tank, it can become sticky and affect normal water addition. Therefore, the water tank and heating water tank must be cleaned every half month and replaced with purified water.