The true density is determined by a gas density meter. We use Accupyc II 1340 and Accupyc II 1345 for measurement, which uses gas displacement technology to determine the volume of the tested sample. Calculate density using the sample weight determined by a balance. This instrument measures the volume of the sample, excluding any voids in bulk powder and any openings in individual particles that allow gas to enter. The internal (closed) porosity is still included in the volume.
This test is usually conducted at room temperature, but it can also be conducted at temperatures specified by the customer within the range of 0C to 50C. Almost any solid sample and some fluid samples can be measured using this technology.
| temperature |
Stable between 15 to 35 ° C (59 to 96 ° F). AccuPyc for temperature control: Temperature stability depends on the specifications of the installed circulation pump. Recommended range: 15 to 50 ° C (59 to 122 ° F) |
| humidity | Relative 20% to 80% (non condensing) |
| height |
14.6 centimeters (5.8 inches) 17.9 centimeters (7.0 inches) for analysis modules with 3 units of 1 centimeter, 10 centimeters, and 100 centimeters 25.9 cm (10.2 inches), 350 cm 3 analysis module 43.0 centimeters (17.0 inches), 2000 cm ³ analysis module |
| width |
27.3 centimeters (10.7 inches), control module (without chamber) 27.3 centimeters (10.7 inches), control module (with compartment) 27.0 centimeters (10.6 inches), 2000 cm ³ analysis module |
| depth | 36.2 centimeters (14.3 inches) |
| weight |
9.3 kilograms (20.5 pounds), control/analysis device (Units of 1, 10, and 100 cubic centimeters) 7.9 kilograms (17.4 pounds), analysis module (Units of 1, 10, and 100 cubic centimeters) 10.5 kilograms (23.2 pounds), analysis module (350 cm ³ unit) 3.6 kg (8.0 lb), control module 26.0 kilograms (57.0 pounds), 2000 cm ³ analysis module |
| voltage | 90 to 264 VAC |
| power | 30 volt ampere |
| frequency | 50 to 60 Hz |
Suggest using research grade helium gas.If not available, please use helium gas with a dew point of -67 ° C (-88 ° F) or lower.Carbon dioxide, argon, dry air, or nitrogen can also be used for different applications (multi gas options can be used to connect multiple gases).
| A 1cm ³ chamber |
1.14 cm ID x 1.1 cm D (0.45英寸 ID x 0.44 英寸D) |
| 10cm ³ chamber |
Inner diameter 1.85 cm x diameter 3.95 cm (Inner diameter 0.72 inches)X diameter1.55 inches) |
| A 100cm ³ darkroom |
Inner diameter 4.62 cm x diameter 6.18 cm (Inner diameter 1.82 inches)X diameter2.43 inches) |
| 350cm ³ darkroom |
5.84cm ID x 13.94cm D (2.30英寸 ID x 5.49 英寸D) |
| A 2000cm ³ chamber |
Inner diameter 9.52 cm x diameter 26.00 cm (Inner diameter 3.80 inches)X diameter10.20 inches) |
| precise | The reproducibility is usually within ± 0.01% of the nominal full-scale sample chamber volume.Using helium gas at 15 to 35 ° C, ensure reproducibility within ± 0.02% of the nominal full-scale volume on clean, dry, and thermally balanced samples. |
| power | Accurate to within 0.03% of the reading, plus 0.03% of the sample capacity |