E-mail
jihpump@163.com
- Phone
-
Address
4F, Building 6, Jinguan Industrial Center, 1076 Konggang Avenue, Yubei District, Chongqing
Chongqing Jieheng Peristaltic Pump Co., Ltd
jihpump@163.com
4F, Building 6, Jinguan Industrial Center, 1076 Konggang Avenue, Yubei District, Chongqing
| Flow Schedule (ml/min, Test Medium: Clear Water, Speed) | ||||||||||
|
Pump head model |
153Yx | 253Yx | ||||||||
| Hose model | 14# | 16# | 25# | 17# | 18# | 15# | 24# | 35# |
36# |
|
|
inside diameterWall thickness |
1.6×1.6 |
3.2×1.6 | 4.8×1.6 |
6.4×1.6 |
7.9×1.6 |
4.8×2.4 |
6.4×2.4 |
7.9×2.4 |
9.7×2.4 |
|
| Reference traffic |
0.03~125 |
0.08~470 |
0.17~1120 |
0.3~1780 |
0.4~2200 |
0.2~900 |
0.3~1620 |
0.4~2340 |
0.5~3000 |
|
| pay attention to |
The flow data in the above table was measured using clean water at room temperature in Jieheng Laboratory and is for reference only 2. The material of the hose, the tolerance of the hose size, the viscosity of the conveying medium, the suction and head can all affect the flow rate 3. When transporting different liquids, it is necessary to consider the compatibility between the liquid and the hose. Please provide specific detailsclickreference resources:Chemical Compatibility Table 4. The lifespan requirements for hoses vary depending on the application scenario. Please refer to the specific requirementsclickreference resources:Hose life table |
|||||||||
| Flow Schedule (ml/min, Test Medium: Clear Water, Speed) | ||||||||
| Pump head model | DGX | 314D | ||||||
| Hose model | 14# | 16# | 25# | 14# | 16# | 25# | 17# | 18# |
| Inner diameter x wall thickness | 1.6×1.6 | 3.2×1.6 | 4.8×1.6 | 1.6×1.6 | 3.2×1.6 | 4.8×1.6 | 6.4×1.6 | 7.9×1.6 |
| Reference traffic | 0-150 | 0-450 | 0-593 | 0.03-162 | 0.08-630 | 0.17-1056 | 0.3-1848 | 0.4-2280 |
| pay attention to |
The flow data in the above table was measured using clean water at room temperature in Jieheng Laboratory and is for reference only 2. The material of the hose, the tolerance of the hose size, the viscosity of the conveying medium, the suction and head can all affect the flow rate 3. When transporting different liquids, it is necessary to consider the compatibility between the liquid and the hose. Please provide specific detailsclickreference resources:Chemical Compatibility Table 4. The lifespan requirements for hoses vary depending on the application scenario. Please refer to the specific requirementsclickreference resources:Hose life table |
|||||||
| Display mode | True color LCD graphics and text |
| Speed range | 0.1~600.0rpm, Reversible reversal |
| Speed regulation method | Rotating encoding switch |
| Operation mode | Flow rate, timing, filling |
| Basic functions | Full speed emptying, suction angle setting, flow calibration, time adjustment, environmental temperature detection |
| Retraction angle | 0~360 ° arbitrary setting |
| Foot switch | Automatic recognition (can be set to jog or run continuously) |
| Filling operation | Can be set to run automatically or with infrared sensing (bottle shortage and stop filling) |
| External dimensions | Length 304 x Width 166 x Height 178 (mm), overall weight: 4.1-5.1Kg |
| Working environment | 0-60 ℃, relative humidity less than 85% |
| Power supply power | Ac220V50/60Hz (Ac110V50/60Hz optional), consumes 54W power |
| Enduring pressure | ≤0.17Mpa, Short term up to 0.27MP (hose outlet back pressure) |
| External control method | Start stop, direction, speed control: 0~5V, 0~10V, 4~20mA, MODBUS_TCP; RS485 communication adopts MODBUS universal protocol |
| Protection level | IP31 |