FBM-160/FBM-100A fluoride ion concentration analyzer

NegotiableUpdate on 06/30
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Using the principle of ion selective electrode, measure the corresponding potential difference for fluoride ion concentration in water. The sensor can be directly immersed in the water sample to measure the concentration of fluoride ions. Application industry: mainly used for online monitoring of fluoride ion concentration in industrial process water and wastewater treatment processes. FBM-160/FBM-100A fluoride ion concentration analyzer

Product Details

FBM-160/FBM-100A fluoride ion concentration analyzer

Product features:

Using the ISE (ion selective electrode) principle, it has good selectivity for fluoride ion measurement and can replace the electrode membrane

There are multiple installation methods for sensor sheaths to meet the requirements of different occasions

The rainproof structure design of the controller can be installed on-site through pipelines or wall mounting. Easy to install, easy to set up and operate

Fluorine ion composite electrode, does not consume reagentsFBM-160/FBM-100A fluoride ion concentration analyzer

● Adopting unfilled reference electrode and porous fluorocarbon resin for salt bridge, with strong anti pollution ability

The instrument has a self diagnostic function

● Can set 4-point alarm output

TECHNICAL INDEX
measuring range 0~99.9mg/L,0~999mg/L,0~9990mg/L
Temperature output 0.0~50.0 ℃ (FBM-160 only)
Instrument characteristics Linear: ± 8% FS (controller)
Reproducibility: ± 5% FS (controller)
90% response time: Within 60 seconds
Analog Output DC 4~20mA (impedance 650 Ω, isolated output)
Analog output range The measurement range above 1/10FS can be arbitrarily set
Factory settings: 0.0~20.0 mgl/L
0~200 mgl/L
0~2000 mgl/L
Digital Output RS-232C (optional)
ALARM IN 4 points can be set
Power requirements AC 90~264V,50/60Hz
environmental condition Outdoor installation; Temperature and humidity: -20~55 ℃, below 95% RH
Installation method FBM-160:50A pipeline or wall mounted or bracket installation
FBM-100A: Indoor panel installation, DIN 92
protection grade FBM-160:IP65 (FBM-100A:IP30)
Sample conditions pH: 4~9pH
temperature 0~40 ℃ (no icing)
conductivity Zui 50ms/m (500 μ S/cm)
velocity of flow 0.01~0.2m/S
Coexisting components

Does not contain a large amount of substances such as calcium, aluminum, iron, etc

Controller model

FBM160-0-AAA0AA0B (0.0~99.9mg/L, wiring G1/2, 50A)

FBM160-0-BA0AA0B (0~999mg/L, wiring G1/2, 50A)

FBM160-0-CA0AA0B (0~9999mg/L, wiring G1/2, 50A)

FBM100A-0-A0B (0.0~99.9mg/L, wiring G1/2)

FBM100A-0-B0B (0~999mg/L, wiring G1/2)

FBM100A-0-C0B<(0~9999mg/L, wiring G1/2)

Note: The above models do not include RS-232. If you have any other requirements, please inquire separately.

Immersion installation - electrode bracket

HCD70C-3-400A010 (PVC material, length 2m)

HCD70F-3-400A010 (PVDF material, 2m long)

HCD76-3-400A010 (PP material, length 2m)

Note: The bracket length is available from 0.5 to 4 meters. Please inquire about the model separately

Circulating installation - circulating pool

HCD82-0-1B (SUS316 material, Rc1/2 connection)

HCD86-1-11B (PP material, Rc1/2 connection)

HCD86-1-21B (PVC material, Rc1/2 connection)

Fluoride ion selective electrode

ELCP-81-5F (5-meter cable) ELCP-81-10F (10 meter cable)

spare parts

Electrode replacement film: 7208L

Fluoride ion standard solution (F-1000mg/L, 500mL): 143F077 ion strength

Regulator (pH5-AB, 500mL): 143A053