FMS portable animal respiratory metabolism measurement system
Product Details

FMSThe portable box integrates three parameter gas analysis units (including water vapor analyzer, CO2 analyzer, and O2 analyzer), gas sampling unit, data collector, and Baseline module (dual channel gas path converter) for precise measurement of animal respiration and energy metabolism. Its main technical features are as follows:

便携式动物呼吸.png

1)Equipped with a baseline unit, it can manually or automatically switch between measuring atmospheric CO2, O2, and other contents (baseline) and indoor CO2, O2, and other contents in the respiratory chamber, thereby more accurately measuring and monitoring animal oxygen consumption rate and CO2 emission rate;

2)The gas sampling module includes a diaphragm pump, a flow control valve, and an air flow meter, which precisely regulate the air flow velocity to accurately measure gas flux through closed or open methods;

3)The gas analysis module includes water vapor analysis module, CO2 and O2 analysis module, built-in temperature and atmospheric pressure sensors, automatic temperature and pressure compensation, high stability and accuracy, and can achieve a resolution of up to 0.1ppm; Optional external methane analyzer;

4)Quick response, gas analysis response time is less than 1 second, which can reflect instantaneous changes in breathing in real time;

5)Dual LCD display screen (with backlight), control and display air humidity RH, dew point temperature, CO2 content, O2 content, flow rate, etc. online through Mode, Adjust, Enter control keys

6)14Channel data collector, capable of connecting 6 temperature sensors and other sensors, with online data storage function;

7)Multi channel measurement experiments can be conducted by selecting an 8-channel gas path converter (connecting multiple animal breathing chambers)

FMS图片.jpg

TECHNICAL INDEX

1.Oxygen Measurement Analysis: Fuel Cell O2Analyzer, unaffected by water vapor, CO2 and other gases, measuring range 1-100%, resolution 0.001%, low noise and high stability, accuracy 0.1%, drift less than 0.02% per hour at constant temperature;

2.Carbon dioxide measurement analysis: Dual band non dispersive infrared technology, measurement range 0-10%, resolution 0.0001% (up to 0.1ppm), accuracy 1%, drift less than 0.001% per hour under constant temperature and oxygen, response time less than 1 second;

3.Water vapor measurement analysis: thin film capacitive sensor, measuring unit is relative humidity or dew point temperature or water vapor partial pressure, measuring range is 0-100% RH, resolution is 0.001% RH, dew point temperature is 0.01 ℃, accuracy is 1%, drift under constant temperature is less than 0.01% per hour

4.CH4Analyzer (external alternative): Dual band non dispersive infrared technology, range 0-10%, accuracy better than 1%, resolution 1 ppm/0.0001%, drift below 0.002% at constant temperature, with LCD display screen and operation keys;

5.Built in airflow pump: anodized aluminum, ball motor (low noise, stable), airflow of 10-2000ml/minute;

6.Airflow control: Micro electronic thermal feedback system, airflow control is actually connected to the airflow pump and flow meter (microcomputer controlled) through a precision feedback loop system, while providing high-precision needle valves;

7.Airflow accuracy of 2%, resolution of 0.1ml/min;

8.The standard range for atmospheric pressure measurement is set to 0-125kPa, and users can set it to 75-125kPa or 90-110kPa according to their needs. The accuracy is 1% and the resolution is 1Pa,

image.png

9.Data acquisition system: 14 simulation channels, data acquisition recording interval of 0.1 seconds to 1 hour, a total of 8000 data points (several hours) can be recorded and stored, and the data can be downloaded from the computer within a few seconds;

10.Thermistor probe is used to measure the temperature inside the breathing room and the air temperature: measurement range -5-60 ℃, resolution 0.001 ℃, absolute accuracy 0.2 ℃, BNC connection, probe diameter 2.5mm;

11.Power supply: 12-15 VDC, optional portable rechargeable battery, AC power, 40Ah battery can work continuously for more than 5 hours in the field

12.Optional military grade waterproof, dustproof, and shock resistant field laptop (with GPS) for field system settings, data browsing, downloading, and analysis, N450 1.66GHz, 8.4 "sun readable LCD display screen, touch screen, 2GB RAM, 80GB SSD

13.The software can display and analyze data online

14.Built in airflow control sampling unit, suitable for small animals such as insects and small rodents; Larger animals such as rodents and rabbits require an external airflow controller MFS-5, including an airflow pump and flow meter, with a flow rate of 250ml/min-5 liters/minute; Large animals such as cows and sheep need to be equipped with an external high flow airflow controller Flowkit 500 (or Flowkit 1000), with a flow rate of 50-500L/min, a resolution of 1L/min, and an accuracy of 2% of the full range;

15.Host size: 48 x 37 x 18 cm;

16.Weight: 7 kg

Origin: United States