Welcome Customer !

Membership

Help

Tengzhou Hongxiang Analytical Instrument Co., Ltd
Custom manufacturer

Main Products:

pharmamach>Products

Tengzhou Hongxiang Analytical Instrument Co., Ltd

  • E-mail

  • Phone

    15163238008

  • Address

    Shandong 1368 Tengfei East Road, Tengzhou Economic Development Zone

Contact Now

Gas chromatograph GC-7890

NegotiableUpdate on 12/30
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Gas chromatograph GC-7890

Product Details

Product Introduction:

GC2080特点.jpg
GC-7890 main technical specifications:
●Operation display:192×64Dot matrix LCDDisplay in both Chinese and English.

● Temperature control range: room temperature+5℃~400℃, incremental:1℃, accuracy:±0.1℃

● Program heating order:16step Cheng Sheng rate:0.1~40℃/min

●detectorcategory:FIDTheTCDTheECDTheFPDTheNPD 5Seed detectorcanselect
●Communication interface:Ethernet:IEEE802.3

●Host size:660×560×480(mm)


Technical parameters of GC-7890 detector:

Detector:/Indicator:

Sensitivity/Detection Limit

Baseline noisesound

Baseline drift

linear range

radioactive source

Hydrogen flame ionization detector (FID)

≤3×10-12g/s
(Hexadecane)/Isooctane)

≤5×10-14A

≤1×10-13A/30min

≥107


Thermal conductivity detector (TCD)

S≥10000mV•ml/mg(benzene/Toluene)
(
magnify1、2、4、8Multiple Choice)

≤20μV

≤30μV/30min

≥105


Electron Capture Detector (ECD)

≤1×10-14g/ml
(C body 666)/Isooctane)

≤0.03mV

≤0.2mV/30min

≥104

63Ni

Flame photometric detector (FPD)

(S)≤2×10-11g/s
(
P)≤1×10-12g/s
(
Methyl parathion/Anhydrous ethanol)

≤3×10-13A

≤2×10-12A/30min

Regarding sulfur≥103
Regarding phosphorus≥104


Nitrogen phosphorus NFD
≤1×10-12G/s (azobenzene)
≤1×10-12g/s (Malathion)

≤Satisfaction rate 3%
≥102
≥103