The honeycomb cell culture system is a specialized equipment for cell product cultivation and a closed integrated operating system that meets GMP sterile production requirements, providing a highly intelligent and continuous sterile culture environment for cell products. The honeycomb cell culture system is equipped with a combination of transport carts and honeycomb cell culture stations, which can meet the requirements of large-scale and diverse cell culture.
Product Details

Product Usage

Suitable for the cultivation of the vast majority of engineered cell products on the market, applied to large hospitals, cell research institutions, cell therapy companies, etc.


Product Features

1. Provide a Class A clean environment for the entire process of cell production and cultivation, in compliance with GMP sterile requirements.

2. Real time recording and storage of system control and environmental monitoring data, achieving data traceability throughout the entire production and cultivation process.

3. Compact integrated design effectively saves space without the need for large building layouts.

4. Just need to install on D-levelandabovecleanIn a clean environment, the construction and operation costs of high-end cleanrooms have been reduced.

5. The honeycomb cultivation system adopts modular design, which can flexibly expand the cultivation space and meet the requirements of large-scale and simultaneous cultivation of different batches of cells.

6. The visual cultivation process monitoring system can identify each independent cultivation unit and manage batches through monitoring, meeting the electronic record, electronic signature, and audit tracking requirements of FDA 21 CFR part 11.

7. Provide a highly secure sterile cell culture environment and utilize hydrogen peroxide sterilization technology to achieve multidimensional sterilization protection for the incubator.

8. The integrated rapid sterilization station can sterilize the cultivation unit separately.

9. Equipped with a transport cart to achieve fast and flexible docking between the incubator and the cell preparation workstation, ensuring that the entire process of cell operation and cultivation is in a sterile environment and preventing cross contamination.


Meets professional standards

1. Complies with the requirements of the appendix for Chinese GMP cell therapy products

2. Meet the production requirements of Chinese GMP sterile products

3. Comply with the technical guidelines for research and evaluation of Chinese cell therapy products

4. Meet the requirements of relevant regulations and industry guidelines (ISO, PDA, USP, and Chinese Pharmacopoeia) of various regulatory agencies (FDA, EMA, NMPA)


Technical Specifications

Honeycomb cultivation system

Internal volume of incubator chamber

30L、60L、100L

Temperature setting range

RT+550℃

temperature resolution

0.1℃

Temperature display error

± 0.1 ℃ (at 37 ℃)

Temperature control error

±0.2℃ (at 37 ℃)

Temperature control fluctuation

±0.2℃ (at 37 ℃)

Temperature control uniformity

±0.3℃ (at 37 ℃)

Thermal insulation performance (secondary temperature difference)

≤8℃

relative humidity

90%RH

power supply

AC220V

CO2Concentration control range

020%

CO2Concentration display resolution

0.1%

CO2Concentration display error

± 0.1% (at a concentration of 5%)

CO2Concentration control error

±0.3%(At a concentration of 5%)

Single incubator power

200-350W

Cabin leakage rate

0.5%Vol/h, Test pressure 100PaUnder the given conditions

Noise during work

≤65dB(A)

data processing

Real time data recording, storage can be transmitted with honeycomb data

alarm function

a. Alarm for high/low temperature; b. Alarm for high/low concentration of carbon dioxide; c. Water shortage alarm for water pan; d. Temperature sensor guarantee, carbon dioxide concentration sensor failure, fan failure, etc