Thermosel heater viscosity measurement range cP (mPa · S)
| SPINDLESAMPLE VOLUMESHEAR RATE(sec-')+MODEL |
SC4-18 8mL 1.32N |
SC4-31 10mL .34N |
SC4-34 9.5mL .28N |
SC4-21 8ml .93N |
SC4-27* 10.5mL .34N |
SC4-28 11.5mL .28N |
SC4-29 13mL .25N |
HT-DIN-81** 7mL 1.29N |
| DVNXLV |
1.2-30K |
12-300K |
24-600K |
Not applicable for historical reasons. However, it is possibleto use the above spindles with any of these instruments3.4-10Kto use the above spindles with any of these instrumentsif you require information on viscosity range. |
1.0-10K |
| DV2TLV |
1.5-30K |
15-300K |
30-600K |
3.4-10K |
| Dvpllv |
1.5-30K |
15-300K |
30-600K |
3.4-10K |
| LVT |
5-10K |
50-100K |
100-200K |
5.7-10K |
| DVNXRV |
However, it is possible to use the abovespindles with any of these instruments. Digital Viscometers/Rheometers willautomatically calculate viscosity. Pleaseyou require information on viscosity range. |
20-500K |
100-2.5M |
200-5M |
400-10M |
14.6-10K |
| DV2TRV |
25-500K |
125-2.5M |
250-5M |
500-10M |
36.5-10K |
| DVPLRV |
25-500K |
125-2.5M |
250-5M |
500-10M |
36.5-10K |
| RVT |
50-100K |
250-500K |
500-1M |
1K-2M |
36.5-10K |
| DVNXHA |
40-1M |
200-5M |
400-10M |
800-20M |
29.2-10K |
| DV2THA |
50-1M |
250-5M |
500-10M |
1K-20M |
73.0-10K |
| DVPLHA |
50-1M |
250-5M |
500-10M |
1K-20M |
73.0-10K |
| HAT |
100-200K |
500-1M |
1K-2M |
2K-4M |
73.0-10K |
| DVNXHB |
160-4M |
800-20M |
1.6K-40M |
3.2K-80M |
116.8-10K |
| DV2THB |
200-4M |
1K-20M |
2K-40M |
4K-80M |
292.0-10K |
| DVPLHB |
200-4M |
1K-20M |
2K-40M |
4K-80M |
292.0-10K |
| HBT |
400-800K |
2K-4M |
8K-16M |
4K-8M |
292.0-10K |
M=1 milionK=1t
housandN= RPMtSpindle SC4-18 1.32x 10 (rpm)= 13.2 sec-1 cP= Centipoise mPa*s = miliPascalseconds
*100 units of disposable rotor SC4-27D can be selected, order number: SC4-27D-100, requiring a special joint adapter, order number: SC4-DSY.
**81 rotor, order number: HT-DIN-81, suitable for HT-2 or HT-2DB sample cups.
Several factors that contribute to stabilizing the testing environment:
The temperature control fluctuation is small.
The sample size is small, and the insulation performance of the sample cup is good, which can effectively reduce the phenomenon of uneven sample temperature
The rotating rotor plays a role in stirring.
The testing process steps are simple and clear. After becoming familiar with the instrument, users can easily obtain accurate and reproducible data without much skill.
Features and advantages
The sample heating control can reach a maximum temperature of+300 ° C
Programmable controllers can provide single point set points or up to 10 temperature point settings. The RheocalcT software can be used to set the temperature gradient between points
Attention: When using RheocalcT software for quality control, it is required to select HT-106 cable.
Can be used with Brookfield standard viscometer and DVNext rheometer
Attention: When used in conjunction with the DVNext rheometer, it is required to select the DVP-141 cable.
Equipped with reusable stainless steel sample holders and five disposable aluminum sample cups
Output: 0-4 VDC suitable for chart recorders; RS-232 is suitable for use with DVNext and/or RheocalcT software.