Verification of Compressed Air System in Medical Device Cleanroom
1、 Compressed air for medical devicesDQTheIQTheOQThePQ
Design confirmation(DQ)Verify whether the system design complies with regulations and process requirements
Installation confirmation(IQ)Verify that the equipment installation meets the design requirements (such as pipeline welding quality, filter model, pressure gauge calibration status).
Operation confirmation(OQ)Test the stability of the system under different operating conditions (such as pressure fluctuation range, dew point compliance, filter pressure difference).
Performance Confirmation(PQ)Under continuous production conditions, detect the indicators of particles, microorganisms, oil content, and moisture at the gas point
Verification of Compressed Air System in Medical Device Cleanroom Compressed air detection-Compressed gas oil-water testCompressed gas detection Oil and water, dew point testing, solid particulate matter (dust particles), microorganisms (planktonic bacteria testing), Suzhou Yikang Environmental Testing uses international brand testing equipment to provide you with accurate testing results.
serial number |
Detect product category |
Product or parameter name |
According to the standard of the subject matter(method)serial number |
Based on the standard(method)name |
1 |
compressed gas |
humidity |
GB/T13277.3-2015 |
Compressed air3Part: Humidity measurement methods |
2 |
compressed air |
oil content |
GB/T 13277.2-2015 |
Compressed air2Part: Measurement method for suspended oil content |
3 |
compressed gas |
solid particles |
GB/T 13277.4-2015 |
Compressed air4Part: Solid particle measurement methods |
4 |
compressed air |
Active microorganisms |
GB/T 13277.7-2021 |
Compressed air7Part: Methods for measuring the content of active microorganisms |
Compressed air water detection |
Oil content and testing—Using GermanyDrager(Delger)ALPHACompressed air quality tester, capable of detecting3~15Compressed gas with a pressure range of kilograms, detection time is only5minute. The side wall of the moisture detection tube is marked with3Column scale and corresponding detection time, observe the numerical scale corresponding to the detection time during detection, and observe the changes in the detection value within the corresponding time (i.e. color changing display) should not exceed the maximum limit of the column value |
Measurement point moisture content={measured value/Detection Time} *Scale standard time |
|||||||||||||||||
| |||||||||||||||||||
|
compressed airWater content(dew point)Testing |
|
||||||||||||||||||
Compressed air oil detection |
Moisture detection—Using GermanyDrager(Delger)ALPHACompressed air quality tester, capable of detecting3~15Compressed gas with a pressure range of kilograms, detection time is only5minute. |
Pull out the oil detection box and observe And compare with the standard view to determine the detection value. |
|||||||||||||||||
Reference diagram for compressed air oil detection |
|
||||||||||||||||||
| |||||||||||||||||||