The walk-in light test chamber is used to investigate the laws of drug changes over time under the influence of temperature and light, providing scientific basis for the production, packaging, storage, and transportation conditions of drugs. At the same time, it establishes the expiration date of drugs through experiments and is suitable for testing the influencing factors, acceleration tests, and long-term tests of drugs;
The walk-in light test chamber is used to investigate the laws of drug changes over time under the influence of temperature and light, providing scientific basis for the production, packaging, storage, and transportation conditions of drugs. At the same time, it establishes the expiration date of drugs through experiments and is suitable for testing the influencing factors, acceleration tests, and long-term tests of drugs.
Technical Specifications:
1. Temperature range: 10-65 ℃
2. Temperature fluctuation: ≤± 0.5 ℃
3. Temperature uniformity: ≤ 2 ℃
4. Range of illumination: 0-100000 LUX
5. Illuminance deviation: ≤± 500LUX
6. UV irradiance range: 0~50W/㎡
7. UV illumination deviation: ± 15%
8. The time required for ICH guidance recommendations of 1.2 million lux · hr is 200 hours
9. Time required for ICH guidance recommendation of 200W · hr/㎡:=50Hr
10. Controller: Color touch screen controller
11. Temperature/humidity sensor: PT100 platinum resistance temperature sensor
12. Refrigeration system: dual sets of fully enclosed high-temperature resistant compressors
13. Data output: dedicated micro printer
14. Inner cavity material: stainless steel (SUS304)
15. Outer box material: Steel plate (SECC) spray painted
16. Protection system: Overtemperature protection, Overheating protection, Overcurrent protection, Overvoltage protection, Grounding protection
17. Power supply: AC220V/50Hz
18. Installation power: determined according to customer specifications
19. Capacity: Customized according to customer specifications and sizes
The walk-in illumination test chamber shall comply with the requirements of the new 2020 edition of the Pharmacopoeia and be manufactured in accordance with the relevant provisions of GB/T10586-2006.

Service commitment:
1. Installation and debugging:
-After the equipment arrives at your company, we will dispatch after-sales service engineers to your company for installation and commissioning within the time required by your company, and all necessary expenses will be borne by us.
2. Operation training:
After the installation and debugging of the equipment are completed and the acceptance is qualified, our technical engineers will provide on-site explanations and demonstration training to the relevant personnel of your company. The demand side will provide training venues and necessary conditions. The training includes:
-Operation Training - Through training, your company's relevant personnel will become proficient in the entire process of equipment operation and understand the principles of the equipment.
-Maintenance Training - Through training, your company's relevant personnel will master the working principles of the equipment, the structure of the whole machine and accessories, and general troubleshooting methods.
-Maintenance Training - Through training, your company's relevant personnel will understand the working principle of the machine, as well as the daily maintenance methods of accessories and the entire machine.
3. After sales service standards:
-From the date of delivery, the warranty period for our company's entire equipment is one year
-During the warranty period, if there are any quality issues with the equipment, our company is responsible for free repair and replacement of parts, without any cost to the customer.
-Our company provides free upgrade services for the control part and software system of the equipment, as well as technical support for the hardware system.
-For specific operators of customers, our company provides free on-site training, simple troubleshooting, and in-depth technical development to ensure that equipment users can master the operation.
-During the warranty period, the after-sales service department shall provide at least two free telephone follow-up visits per year (including machine maintenance, technical explanations, and operation guidance)