- Phone
-
Address
B3 Floor, Wuhan University Science and Technology Park, Daxueyuan Road, Donghu New Technology Development Zone, Wuhan City
Wuhan Zhongyi Baikong Instrument Co., Ltd
B3 Floor, Wuhan University Science and Technology Park, Daxueyuan Road, Donghu New Technology Development Zone, Wuhan City




The working principle of an ultrasonic level gauge is that the transducer (probe) emits ultrasonic pulses that are reflected back from the surface of the measured medium. Some of the reflected echoes are received by the same transducer and converted into electrical signals. Ultrasonic pulses propagate at the speed of sound waves, and the time interval required from emission to reception of ultrasonic pulses is proportional to the distance from the transducer to the surface of the measured medium. This distance valueThe relationship between S, sound velocity C, and transmission time T can be expressed by the formula: S=CxT/2.

2. After the equipment installation is completed, it must be truly grounded separately and not grounded with the common ground of the electrical box or instrument box.
3. Suggestion: When connecting ultrasonic level gauges to equipment that interferes with frequency converters, PLCs, etc., isolation transformers should be added to the power supply, signal isolators should be added to the signal section, and reliable grounding treatment should be carried out.
4. The signal line cannot be installed in the same cable tray as the power line or power line. It should be installed separately in a metal tube, or installed away from the power line or power line. Without installation in a tube, the distance from the power line or power line should be at least 1 meter.