Comprehensive drug stability test boxThere are also requirements during installation. Let's take a look at its installation requirements below:
Comprehensive drug stability test boxInstallation requirements:
1. Environmental conditions
⑴ Temperature: 5-35 ℃;
⑵ Relative humidity: ≤ 85% R.H;
⑶ Atmospheric pressure: 86-106Kpa;
There is no strong vibration or electromagnetic field around;
There is no high concentration of dust or corrosive substances around, and there is no direct sunlight or direct radiation from other heat sources;
The equipment should be placed horizontally in a well ventilated laboratory, with a 1-meter space left around for operation and maintenance
2. Equipment placement
The equipment should be placed on a flat surface with sufficient space around for maintenance and air flow ventilation.
The location of the test chamber should be away from flammable, explosive, volatile, and corrosive materials.
3. Power connection
Power supply: The system power supply adopts a single-phase three wire system, with a voltage of 220 ± 22V, 50Hz, and 10A. When installing power sockets, the grounding wire must not be left floating or disconnected to prevent electric shock accidents.
The above are the installation requirements for the comprehensive drug stability test box.