The compact and sturdy design of the 11-D combines all the advantages of the previous portable GRIMM aerosol spectrometer, improving the performance of the optical detection chamber, longer battery life, and easier operation, making the 11-D a leader in portable aerosol particle size spectrometers. 11-D can detect every particle in real-time, with a particle size range of 0.25 μ m~35 μ m and 31 particle size channels. Users can obtain and verify monitoring data in real-time in the studio through a computer, and in addition, the data can be stored on a memory card.
The measured results are displayed in real-time by counting or mass concentration, and the detected gases can be collected on a 47mm PTFE filter membrane for further chemical, microscopic, or gravity analysis. Using a constant velocity sampling probe, it can collect airflow of 4~25m/s and directly collect the gas before and after filtration from the pipeline in the airway.
Characteristic
Small, portable, durable, battery powered
Longer battery life (10h)
The 47mm PTEF filter membrane can collect samples and provide further gravity analysis and chemical analysis
31 equidistant particle size channels, traceable PSL ball calibration
Multi functional data acquisition and communication interface (Bluetooth, Ethernet, USB, RS232)
The 47mm PTEF filter membrane can collect samples and provide further gravity analysis and chemical analysis
Upon startup, all optical and pneumatic components undergo self inspection to meet high-quality standards
Circulating cleaning gas to ensure the cleanliness of the light detection cell and laser light source
Sampling flow rate of 1.2LPM, constant current control
Real time acquisition of PM10, PM4, PM2.5, PM1.0, PMcoarse through software
The monitoring data output is the particle number concentration, and the mass ratio of particles in the 31 particle size channel is output through software (software: compliant with European standard EN481, occupational health particle concentration)
Theory
Indoor Air Quality
Aerosol research
Pollution source analysis
Workplace assessment
Measurement of particulate matter migration
Monitoring of industrial production sites

Parameter
performance parameters
Detection principle: Single particle 90 ° light scattering (based on laser diode), aerodynamic focusing, full sampling volume without dead angle detection, no sampling boundary error. [μm]
Detector: Ultra fast signal processor, 2 * 16 raw data channels
Time resolution: 6s, 31 channels (optional 6s, 1.5, 10, 15, 30, 60 minutes)/1s, 16 channels (0.253~2.982 μ m, or 2.982~35.15 μ m)
Measurement particle size channels: 0.253~35.15 μ m, 31 equidistant particle size channels
Quantity concentration measurement range: 1-3 000 000 P/L (higher concentrations can be diluted with an optional diluter)
Measurement range of mass concentration: 0.1 μ g/m3~100 mg/m3
Occupational health assessment: measurement of inhalable particulate matter, thoracic particulate matter, and pulmonary particulate matter (European standard, EN481)
Environmental quality assessment: TSP, PM10, PM4, PM2.5, PM1.0, PMcoarse
Repeatability>97% range (Max), according to ISO 21501-1:
Aerosol sampling flow rate: 1.2LPM ± 3% constant control, in accordance with ISO 21501-1
Flushing air flow rate: 0.4LPM constant control, automatic cleaning in standby mode (instrument start and pause modes)
Sample collection filter membrane: 47mm diameter PTFE filter membrane
Laser source: diode laser 683nm (Pmax=40 mW)
Operation: touchpad or PC (software or super terminal)
LC display: 2 x 16 alphanumeric characters
Self check: Automatically perform self check after each startup
Storage interval: 1 minute to 1 hour optional
Specifications
Data storage: standard 2GB USB flash drive
Communication methods: Bluetooth, Ethernet, RS-232 interface, and USB
Data output: LC display, particulate matter mass concentration/number concentration (per minute or sampling volume), warning value, battery level; Particle size factor, measurement location number, date and time; Optional sensor results
Analog input: 1 (0~10V), resolution of 10bits (approximately 10mV)
Battery: Lithium battery 10.8V/6.5Ah, continuous operation for 10 hours, equipped with a speed charging charger
Power adapter: Input: 100~240VAC, 47~60Hz, Output: 13VDC, Maximum current: 2.5A
Size specifications: 27cm × 13cm × 7cm
Weight: 2.1kg
Measurement environment: operating temperature and humidity+4 ℃~+40 ℃ RH<95%、 Non corrosive gas; Storage and transportation temperature and humidity: -20 ℃~+50 ℃, RH<95%; Pressure range:<120hPa, approximately 1000m above sea level
Software
The 11-D dedicated software 1179 can obtain data, measurement results, and instrument functional parameters online, and can be applied on most 32/64 bit Windows systems.
Number concentration: Particle number concentration P/L
Mass concentration: μ g/m3(PM10, PM4, PM2.5, PM1.0, PMcoaleEN481) and the mass concentration of each channel in μ g/m3
Integrated sensors: temperature, humidity, and wind speed sensors
Service data: pump flow, battery level, operational errors, etc