Welcome Customer !

Membership

Help

Hangzhou Gengyu Instrument Co., Ltd
Custom manufacturer

Main Products:

pharmamach>Products

2000ml low-temperature constant temperature stirring reaction bath/reaction tank

NegotiableUpdate on 01/19
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
DHJF-8005 2000ml low-temperature constant temperature stirring reaction bath can replace dry ice and liquid nitrogen for low-temperature reactions. It is mainly used to cool reactors such as flasks, beakers, and test tubes, and is suitable for hospitals, scientific research, universities, and enterprise laboratories.
Product Details

2000ml low-temperature constant temperature stirring reaction bathPerformance characteristics

1. Fast cooling, reaching the temperature you need faster.

2. Temperature digital setting, LCD display.

3. Fully enclosed magnetic stirrer to maintain mixing stability.

4. The diameter of the retractable cover can be adjusted according to the outer diameter of the reaction vessel to reduce refrigerant consumption.

5. Equipped with a built-in secondary stirrer, it works together with the stirrer inside the reaction vessel to maintain a uniform temperature of the sample.

6. Equipped with fixed rods and deep bucket supports, it is convenient to place and fix other supporting devices such as burettes and sensors.

7. Adopting imported compressors, with high reliability and efficiency.

8. There is a liquid outlet for staff to easily replace the shower gel.

2000ml low-temperature constant temperature stirring reaction bathTechnical Specifications

model

DHJF-8005

Storage tank capacity (L)

5

Empty load maximum low temperature (℃)

-80

work

make, do, create, compose, write, etc.

article

document

Temperature range for use (℃)

-80~99

Environmental temperature (℃)

≤25

Relative humidity of the environment (%)

≤60

power supply

220V±10% 50Hz

Lord

want

sex

can

and

ministry

document

temperature control

display mode

LCD (Liquid Crystal Display)

Temperature control accuracy (℃)

±0.2

sensor

Pt100

Instrument adjustment range (℃)

±50

Setting method

Set arbitrarily within the temperature range of use, with a resolution of 0.5 ℃

safety protection

Delay, overcurrent, overheating

Cooling capacity (W)

320-1750

refrigeration unit

Power (W)

2830

Current (A)

12.6

refrigerant

R404A/R23

Heating power (W)

1500

Mixing power (W)

30

circulation pump

Power (W)

100

Flow rate (L/min)

20

Head (m)

5~7