Welcome Customer !

Membership

Help

Beijing Baoyun Xingye Science and Trade Co., Ltd
Custom manufacturer

Main Products:

pharmamach>Products

Beijing Baoyun Xingye Science and Trade Co., Ltd

  • E-mail

  • Phone

  • Address

    1606, Block A, Langqin International Building, No. 168 Guang'anmen Outer Street, Xicheng District, Beijing

Contact Now

202 portable ozone detector

NegotiableUpdate on 03/11
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Product Introduction: Model? The 202 type ozone detector is designed specifically for accurately detecting ozone concentration in the air. The instrument uses standard 254nm ultraviolet absorption technology, with a detection range from 0pppb to 250ppm and an accuracy of up to ppb level. The instrument has the characteristics of high accuracy and low loss.
Product Details

    Product Introduction

    The Model 202 ozone detector is designed specifically for accurately detecting ozone concentration in the air. The instrument uses standard 254nm ultraviolet absorption technology, with a detection range from 0pppb to 250ppm and an accuracy of up to ppb level. The instrument has the characteristics of high accuracy and low loss.

    Product Application

    ● Atmospheric measurement of balloons and aviation

    Monitoring in industrial production

    Establish a city monitoring network

    ● Personal and work environment monitoring

    ● Indoor air detection

    Product Features

    ● Built in long-life sampling pump

    ● Low energy consumption

    Internal data logger

    Real time time display

    Eliminate any water vapor interference

    ● Large capacity memory card, providing mobile data storage

    ● Serial interface USB connection port


    Product Parameters

    Detection Principle

    UV absorption, 254nm

    Testing Scope

    0 ppb ~ 250 ppm

    precision

    0.1ppb

    error

    1.5ppb or 2% of the reading, whichever is higher (10s test mode)

    threshold value

    3.0 ppb

    NIST traceable calibration

    have

    Measurement interval

    10 seconds (average data options: 10 seconds, 1 minute, 5 minutes, 1 hour)

    baseline drift

    <2 ppb/day,<5 ppb/year

    Traffic (normal)

    1L/min

    Traffic demand

    > 0.6L/min

    Sensitivity drift

    <1%/day,<3%/year

    Measure time and frequency

    10s,0.1Hz

    response time

    20s, two o'clock

    Operating temperature range

    Standard: 0-50 ℃, optional: -20~50 ℃ (low temperature mode)

    average time

    10 seconds, 1 minute, 5 minutes, 1 hour

    data storage

    14336 lines (10 s average=1.4 days); 5-minute average=1.4 months)

    data output

    RS232, 0~2.5 V, LCD 显示

    Power supply requirements

    Battery or 110/220V battery pack

    Ozone unit

    ppb、ppm、pphm、μg/ m3、mg/m3

    temperature unit

    ℃、 K

    pressure unit

    Mbar、torr

    Weight/Size

    2.8kg, 1.3kg (excluding instrument box)/9.5 × 21.5 × 30 cm