- Phone
-
Address
Building B, 588 Shengang Road, Songjiang District, Shanghai
Shanghai Moxi Scientific Instrument Co., Ltd
Building B, 588 Shengang Road, Songjiang District, Shanghai
Product Introduction
The binder KBF series constant temperature and humidity chamber can provide a stable temperature and humidity environment, simulate various climatic conditions, and is particularly suitable for stability experiments, long-term experiments, equilibrium experiments, and bacterial culture of samples. In durability and stability tests conducted according to internationally recognized EN standards and ICH standards (CPMP/ICH/380/95/Q1A and Q1B), it also meets the standards.

Binding Core Technology - APT.Line@Internal preheating technology enables the temperature inside the cavity to reach a very uniform distribution, changing the traditional preheating system's difficulty in overcoming the temperature difference between the upper and lower neck bottles.
1) Ensure high control accuracy within the entire available space, and the control accuracy can be reproduced.
2) The temperature distribution is relatively uniform, and the temperature adjustment process is smooth and stable.
3) Due to APT.Line@The air conduction performance ensures precise control of space temperature even under full load conditions.

DCT@Direct cooling technology
KBFTechnical characteristics of series constant temperature and humidity chambers
beltDCT@Electronic controlled refrigeration systemATP.Line@Internal preheating technology
1) Temperature range: -10 ℃ -100 ℃ (no humidity), fluctuation ± 0.1 ℃;
2) Humidity range: relative humidity 10% -90%, fluctuation ± 1.5% rH;
3) MCS controller: 25 stored programs, each program has 100 sections, with a maximum of 500 program segments;
User friendly liquid crystal display screen (LCD)
Ø Easy to read menu guide
Ø Comprehensive electronic chart recorder
Various options for displaying process parameter charts
Ø Real clock
4) Electronic controlled humidification and dehumidification systems;
5) Use tap water (conductivity between 200-500 μ S) and humidify the electrodes;
6) Capacitive humidity sensor, precise humidity measurement, no maintenance required;
7) Compliant with ICH guideline Q1A, suitable for sample stability testing;
8) Automatic defrosting device suitable for long-term operation;
9) Internal glass door;
10) Door mounted condensate collection tray;
11) Environmentally friendly refrigerant R 34a;
12) Independent adjustable temperature safety device, Class 3.1(DIN 12880), Equipped with visual and audible alarms;
13) Access port with silicone plug (diameter 30mm), on the right side;
14) Complete water supply safety connection kit, including inlet hose and drain pipe (total length of 6 meters);
15) RS422 interface, used for communication software APT-COM @ data control system;
16) 2 stainless steel shelves;
KBFTechnical specifications for series constant temperature and humidity chambers
|
model |
KBF 115 |
KBF 720 |
|
External dimensions (W × H × D) (mm) |
834×1022×646 |
1234×1816×867 |
|
Internal cavity size (W × H × D) (mm) |
600×480×400 |
1000×1168×600 |
|
Internal cavity volume (L) |
115 |
700 |
|
Stainless steel shelving (standard/maximum) |
2 / 5 |
2 / 14 |
|
Load per shelf/total load (kg) |
20 / 50 |
45 /120 |
|
Number of doors/Internal glass doors |
1 / 1 |
2 / 2 |
|
Temperature range (without humidity) (° C) |
-10-100 |
-10-100 |
|
Temperature range (with humidity) (° C) |
10-90 |
10-90 |
|
Temperature deviation @ 10 ° C (no humidity) (± ° C) |
0.4 |
0.4 |
|
Temperature deviation @ 37 ° C (no humidity) (± ° C) |
0.3 |
0.4 |
|
Temperature deviation (with humidity) (± ° C) |
1 |
1 |
|
Temperature fluctuation range (during heating) (± ° C) |
0.1 |
0.1 |
|
Temperature fluctuation range (during cooling) (± ° C) |
0.5 |
0.5 |
|
Heating time (to 37 ° C) (min) |
23 |
28 |
|
Cooling time (from room temperature to 10 ° C) (min) |
35 |
35 |
|
Recovery time (30 seconds after opening the door to 37 ° C) (min) |
5 |
5 |
|
Recovery time (30 seconds after opening the door to 50 ° C) (min) |
4 |
4 |
|
Humidity fluctuation range (±% relative humidity) |
1.5 |
1.5 |
|
Power supply (± 10% 50/60 Hz) |
230V/1.7KW |
230V/2.8KW |
|
Net weight (kg) |
115 |
278 |